FDA Adverse Event Malfunction Summary report: N

MATRIXMANDIBLE SHORT CUT PLATE CUTTER

MDR report key: 3041689 · Received April 8, 2013

Report

Report Number
8030965-2013-10785
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
September 29, 2011
Report Date
September 29, 2011
Manufacturer
SYNTHES GMBH
Product Code
HTZ
PMA / PMN Number
EXEMPT
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

SYNTHES IS SUBMITTING THIS REPORT AS A RESULT OF REMEDIATION ACTIVITIES RELATED TO FDA WARNING LETTER DATED FEBRUARY 2012. DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR. A REVIEW OF THE DEVICE HISTORY RECORDS WAS PERFORMED AND NO COMPLAINT RELATED ISSUES WERE FOUND. THE PRODUCT DEVELOPMENT EVALUATION REVEALED THAT THE DEVICE IS IN GOOD CONDITION AND IS UNDAMAGED. THERE ARE SOME SMALL NICKS ON THE EDGES OF THE CUTTING SURFACE BUT WOULD NOT IMPACT THE ABILITY OF THE DEVICE TO FUNCTION PROPERLY. THE RETURNED DEVICE IS IN GOOD CONDITION AND IS FULLY FUNCTIONAL AND CAPABLE OF CUTTING PLATES AS DESIGNED. IT IS CONCLUDED THAT THIS COMPLAINT IS INVALID AS THERE IS NO ISSUE WITH THE RETURNED DEVICE.

Additional Manufacturer Narrative · 1

DEVICE(S) LISTED IN THIS REPORT IS (ARE) USED FOR TREATMENT, NOT DIAGNOSIS. ANY ADDITIONAL INFORMATION RECEIVED REGARDING THIS EVENT AFTER FILING THIS REPORT SHALL BE FILED ON A SUPPLEMENTAL MDR.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN ORIF MANDIBLE FRACTURE THE SURGEON WAS USING THE MATRIX MANDIBLE SHORT CUT PLATE CUTTER TO CUT A PLATE IN HALF, AND THE PLATE CUTTERS BROKE. ALL THE PIECES WERE RETRIEVED. TWO MATRIX MANDIBLE SHORT PLATE CUTTERS WERE RECEIVED, BOTH SAME PART NUMBER AND LOT NUMBER. THIS IS REPORT 2 OF 2 FOR THIS EVENT.

Description of Event or Problem · 1

THIS IS REPORT 2 OF 2 FOR (B)(4).

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144050 MATRIXMANDIBLE SHORT CUT PLATE CUTTER HTZ SYNTHES GMBH 3416296

Patients

Seq Age Sex Outcome Treatment
1