FDA Adverse Event Malfunction Summary report: N

TENDRIL ST

MDR report key: 3041494 · Received April 8, 2013

Report

Report Number
2017865-2013-01694
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
February 22, 2013
Manufacturer
ST JUDE MEDICAL CARDIAC RHYTHM MANAGEMENT DIVISION
Product Code
NVN
PMA / PMN Number
P960013
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER HEALTH CARE PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

ALL INFORMATION PROVIDED BY MANUFACTURER, NO MEDWATCH FORM WAS RECEIVED RELIABILITY LABORATORY TECHNICIAN, (B)(6) - FAILURE (EVENT) OBSERVED DURING ANALYSIS. FINAL ANALYSIS FOUND THAT THE INSULATION WAS ABRADED AT 4.5 CM TO 4.7 CM FROM THE CONNECTOR PIN. THE ABRASION WAS CONSISTENT WITH CONSTANT FRICTION WITH ANOTHER IMPLANTABLE DEVICE.

Description of Event or Problem · 1

THIS REPORT IS TO ADVISE OF AN EVENT OBSERVED DURING ANALYSIS.