HAKIM PROG VALVE INLINE SIPHONGUARD
Report
- Report Number
- 1226348-2013-15071
- Event Type
- Injury
- Date Received
- April 8, 2013
- Date of Event
- March 1, 2013
- Manufacturer
- CODMAN & SHURTLEFF, INC
- Product Code
- JXG
- PMA / PMN Number
- PK992173
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OH, US
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
HISTORICALLY, FOR COMPLAINTS OF THIS NATURE, EXAMINATIONS OF THE VALVES HAVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICES WHICH HAVE INTERFERED WITH THEIR PROGRAMMING AND/OR PRESSURE CONTROL MECHANISMS. THE BIOLOGICAL DEBRIS HAS CAUSED THE RETURNED VALVES TO FAIL THE PROGRAMMING AND/OR PRESSURE TESTS AND APPEARS TO HAVE CAUSED THE DIFFICULTIES EXPERIENCED BY THE CUSTOMERS. REVIEWS OF THE DEVICE HISTORY RECORDS HAVE CONFIRMED THAT THE VALVES HAVE MET SPECIFICATIONS WHEN RELEASED TO STOCK. A FOLLOW UP REPORT WILL BE FILED IF THE INVESTIGATION OF THIS DEVICE REVEALS A RESULT DIFFERENT FROM THAT WHICH IS STATED ABOVE OR IF ANY FURTHER INFORMATION REGARDING THE CAUSE OR MODE OF FAILURE IS MADE AVAILABLE. OTHERWISE, THE COMPLAINT IS CONSIDERED TO BE CLOSED AT THIS TIME.
EXAMINATION OF THE VALVE REVEALED THE PRESENCE OF BIOLOGICAL DEBRIS WITHIN THE DEVICE. THIS BIOLOGICAL DEBRIS CAUSED AN OCCLUSION IN THE DEVICE AND FOR THE RETURNED VALVE TO FAIL THE PROGRAMMING TEST, WHICH APPEARS TO HAVE CAUSED THE DIFFICULTY ENCOUNTERED BY THE CUSTOMER. REVIEW OF THE DEVICE HISTORY RECORD CONFIRMED THE VALVE MET SPECIFICATION REQUIREMENTS WHEN RELEASED TO STOCK. BASED ON THE RESULTS OF THIS INVESTIGATION NO FURTHER ACTION IS REQUIRED. TRENDS WILL BE MONITORED FOR THIS AND SIMILAR COMPLAINTS. AT THE PRESENT TIME THIS COMPLAINT IS CLOSED.
THE SALES REP REPORTED THAT THE SURGEON REMOVED A VALVE AS IT WOULD NOT PROGRAM. HE OBSERVED THAT THE VALVE APPEARED TO HAVE BEEN FLIPPED, WHICH MAY ACCOUNT FOR THE PROGRAMMING PROBLEMS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142702 | HAKIM PROG VALVE INLINE SIPHONGUARD | SHUNT, CENTRAL NERVOUS SYSTEM & COMPS | JXG | CODMAN & SHURTLEFF, INC | CNLCHW |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Required Intervention |