Narratives

Additional Manufacturer Narrative · 1

UPON EVALUATION IT WAS DISCOVERED THAT MULTIPLE INTERNAL COMPONENTS WERE BROKEN/DAMAGED INCLUDING THE NOSE HOUSING, FRONT HOUSING, YOKE, AND LINK WITH FINS. IN ADDITION, THERE WAS CORROSION IN THE PRESS PLUG.

Description of Event or Problem · 1

THE MICRO SAGGITAL SAW WAS RETURNED FOR SERVICE. UPON EVALUATION AT THE MANUFACTURER, IT DISPLAYED A BIAS CURRENT MESSAGE WHEN CONNECTED TO THE CONSOLE SIGNALING A CONDITION OCCURED FROM WHICH THE DEVICE HAS THE POTENTIAL TO RUN WITHOUT USER ACTIVATION. THERE WAS NO PATIENT INVOLVEMENT, AND THERE WERE NO USER INJURIES OR ADVERSE CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143155 MICRO SAGITTAL SAW DRILLS, BURRS, TREPHINES & ACCESSORIES (SIMPLE, POWERED) HBE STRYKER INSTRUMENTS-KALAMAZOO