FDA Adverse Event Malfunction Summary report: N

MINICAP TRANSFER SET

MDR report key: 3041349 · Received April 8, 2013

Report

Report Number
1416980-2013-08721
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
January 1, 2013
Report Date
March 15, 2013
Manufacturer
BAXTER HEALTHCARE - MOUNTAIN HOME
Product Code
KDJ
PMA / PMN Number
K882498
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
SF
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). A REVIEW OF ALL BATCH RECORD DOCUMENTS WAS PERFORMED FOR LOT NUMBER H11G25033 WITH NO ISSUES NOTED DURING THE MANUFACTURING PROCESS. THERE WERE NO DEVIATIONS FROM STANDARD PROCEDURE AND NO EXCEPTIONS RELATED TO THE REPORTED CONDITION WERE NOTED. A VISUAL INSPECTION NOTED CRACKS IN THE TWIST CLAMP AS WELL AS ADDITIONAL CRACKS AND DETERIORATION OF THE SAMPLE. NO LEAK TEST OR CLAMP FUNCTION TEST COULD BE PERFORMED DUE TO THE DETERIORATION OF THE SAMPLE. CLEAR PASSAGE TESTING WAS PERFORMED WITH NO ISSUES NOTED. THE SAMPLE WAS CONFIRMED FOR DAMAGE. THE ROOT CAUSE COULD NOT BE DETERMINED.

Description of Event or Problem · 1

IT WAS REPORTED BY A CUSTOMER THAT THE TWIST CLAMP OF A TRANSFER SET WAS CRACKED. THE PATIENT USES BATH SOAP AND WATER TO CLEAN THE TRANSFER SET. THE SAMPLE WAS RECEIVED BY BAXTER, AND UPON EVALUATION, THE TRANSFER SET WAS FOUND TO HAVE ADDITIONAL, NONCOSMETIC DAMAGE. THERE WAS PATIENT INVOLVEMENT, BUT NO PATIENT INJURY OR MEDICAL INTERVENTION WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144093 MINICAP TRANSFER SET SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE KDJ BAXTER HEALTHCARE - MOUNTAIN HOME H11G25033

Patients

Seq Age Sex Outcome Treatment
1