FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3041317 · Received April 8, 2013

Report

Report Number
2124215-2013-01400
Event Type
Injury
Date Received
April 8, 2013
Date of Event
January 16, 2013
Report Date
February 1, 2013
Manufacturer
CPI PLANT - ST. PAUL
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

TO DATE, THE EXPLANTED LEAD HAS NOT BEEN RECEIVED AT BOSTON SCIENTIFIC. UPON RECEIPT THE LEAD WILL UNDER GO DETAILED LABORATORY ANALYSIS IN AN ATTEMPT TO CONFIRM AND DETERMINE THE ROOT CAUSE OF THIS EVENT.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND SET SCREW MARKS ON THE TERMINAL PIN. IN ADDITION, THE PROXIMAL COIL WAS STRETCHED AND FLUID WAS NOTED IN THE HELIX. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES AS THE LEAD PASSED ALL ELECTRICAL TESTING.

Description of Event or Problem · 1

==

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THE PATIENT WITH THIS RIGHT VENTRICULAR (RV) LEAD EXPERIENCED A SHOCK AFTER A RUN OF VENTRICULAR TACHYCARDIA (VT). DURING A ROUTINE FOLLOW UP IT WAS NOTED THAT THE LEAD EXHIBITED NO CAPTURE FOR AN UNKNOWN DURATION AND HIGH THRESHOLDS. A REPLACEMENT PROCEDURE WAS SCHEDULED. BEFORE THE REPLACEMENT, IT WAS NOTED THAT THRESHOLD TESTING WAS PERFORMED FOR A SECOND TIME AND THE PATIENT EXPERIENCED DIAPHRAGMATIC STIMULATION. A LEAD PERFORATION WAS SUSPECTED. THE LEAD WAS THEN EXPLANTED AND REPLACED. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143249 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI PLANT - ST. PAUL 0696

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R