FDA Adverse Event Injury Summary report: N

TELIGEN

MDR report key: 3041243 · Received April 8, 2013

Report

Report Number
2124215-2013-01918
Event Type
Injury
Date Received
April 8, 2013
Date of Event
October 24, 2012
Report Date
February 1, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THIS ICD WAS THOROUGHLY INSPECTED AND ANALYZED. EXTERNAL VISUAL INSPECTION OF THE DEVICE REVEALED NO ANOMALIES. REVIEW OF THE DEVICE MEMORY CONFIRMED THAT FAULT CODE 1003 WAS RECORDED ON (B)(6) 2012. ALTHOUGH THE DEVICE MEMORY DIAGNOSTICS DEMONSTRATED THAT THE DAILY BATTERY VOLTAGE MEASUREMENTS DISPLAYED AN IRREGULAR PATTERN OF DISCHARGE, THE BATTERY VOLTAGE LEVEL AT EXPLANT (2.96 VOLTS) WAS SUFFICIENT TO ENSURE THERAPY AVAILABILITY/DELIVERY WHILE THE DEVICE WAS IMPLANTED. THE DEVICE CASE WAS THEN OPENED TO FACILITATE ANALYSIS OF THE INTERNAL COMPONENTS. THE BATTERY WAS SEPARATED FROM THE OTHER DEVICE ELECTRONICS AND THE OVERALL CURRENT DRAW OF THE CIRCUITRY WAS MEASURED. A NORMAL CURRENT DRAIN WAS OBSERVED WITHIN THE CIRCUITRY. COLLECTIVELY, THE PATTERN OF IRREGULAR DAILY BATTERY VOLTAGE MEASUREMENTS IN CONJUNCTION WITH NORMAL POWER LEVELS AND DEVICE HYBRID CURRENT DRAW IS CONSISTENT WITH BEHAVIOR OF DEVICES WHERE A LATENT CURRENT LEAKAGE PATH HAS OCCURRED WITHIN THE BATTERY ITSELF, RESULTING IN A PARTIAL DEPLETION OF THE BATTERY.

Additional Manufacturer Narrative · 1

THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION DURING AN IN CLINIC INTERROGATION, A FAULT CODE (1003) FOR MEASURED BATTERY VOLTAGE IS TOO LOW FOR PROJECTED REMAINING CAPACITY HAD OCCURRED. BOSTON SCIENTIFIC TECHNICAL SERVICES (TS) WAS CONSULTED AND RECOMMENDED REPLACEMENT OF THE DEVICE. THE DEVICE WAS REPLACED SIX DAYS LATER, A NEW DEVICE WAS SUCCESSFULLY IMPLANTED. THE DEVICE WAS RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143588 TELIGEN IMPLANTABLE CARDIOVERTER DEFIBRILLATOR LWS GUIDANT CRM CLONMEL IRELAND E110

Patients

Seq Age Sex Outcome Treatment
1 65 YR Hospitalization| L| R 4135| 0158| E110