FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3041236 · Received April 8, 2013

Report

Report Number
2124215-2013-02473
Event Type
Injury
Date Received
April 8, 2013
Date of Event
January 15, 2013
Report Date
February 5, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
PR
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE LEAD IS CURRENTLY UNDERGOING ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD DISPLAYED AN INCREASE IN PACING THRESHOLD. THE PATIENT WAS SEEN FOR A LEAD REVISION PROCEDURE. AT THAT TIME, THE PHYSICIAN CONCLUDED THAT LEAD MAY HAVE SLIGHTLY PERFORATED THE PATIENT'S VENTRICLE. THE LEAD WAS EXPLANTED AND REPLACED. THERE WERE NO ADVERSE PATIENT EFFECTS REPORTED. THE LEAD WAS RETURNED FOR ANALYSIS.

Description of Event or Problem · 1

--

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142580 ENDOTAK RELIANCE NVY CPI - DEL CARIBE 0292

Patients

Seq Age Sex Outcome Treatment
1 66 YR Hospitalization| L| R N140| 4554| 4136| 0292