FDA Adverse Event Malfunction Summary report: N

INSIGNIA

MDR report key: 3041234 · Received April 8, 2013

Report

Report Number
2124215-2013-02652
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
January 9, 2013
Report Date
June 12, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
NVZ
PMA / PMN Number
D970003
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE DEVICE WAS PERFORMED. A VISUAL INSPECTION OF THE DEVICE HEADER AND CASE NOTED NO ANOMALIES. [PIN GAUGE TESTING, DESIGNED TO VERIFY PROPER PORT DIMENSIONS, WAS COMPLETED. EACH PORT MEASURED AS EXPECTED. THE DEVICE WAS THEN EXPOSED TO SIMULATED HEART LOAD CONDITIONS, AND THE PACING AND SENSING FUNCTIONS WERE TESTED. THE DEVICE OPERATED APPROPRIATELY, ACCORDING TO ITS PERFORMANCE SPECIFICATIONS, WITH NO INTERRUPTIONS IN THERAPY OUTPUT OR PROGRAMMER COMMUNICATION AT THE RETURNED PROGRAMMED SETTINGS. A SERIES OF ELECTRICAL TESTS WAS ALSO PERFORMED, AND AGAIN, NORMAL DEVICE FUNCTION WAS OBSERVED.

Additional Manufacturer Narrative · 1

OUR RECORDS INDICATE THIS DEVICE REMAINS IN SERVICE. IF ADDITIONAL INFORMATION IS RECEIVED, THIS REPORT WILL BE UPDATED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PACEMAKER DISPLAYED A LONGEVITY OF ONE YEAR APPROXIMATELY FOUR MONTHS AGO. CURRENTLY, IT IS DISPLAYING LONGEVITY OF THREE YEARS. NO ADVERSE PATIENT EFFECTS WERE REPORTED. ADDITIONAL INFORMATION HAS BEEN REQUESTED; HOWEVER, NO FURTHER INFORMATION IS CURRENTLY AVAILABLE.

Description of Event or Problem · 1

SUBSEQUENTLY, THE DEVICE WAS EXPLANTED AND RETURNED FOR ANALYSIS.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143569 INSIGNIA IMPLANTABLE PULSE GENERATOR NVZ GUIDANT CRM CLONMEL IRELAND 1291

Patients

Seq Age Sex Outcome Treatment
1 68 YR 1291| 1388T| 1270