ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-01262
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- January 11, 2013
- Report Date
- June 6, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
ACCORDING TO AVAILABLE INFORMATION, THIS SYSTEM REMAINS IMPLANTED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS EVENT WILL BE UPDATED.
- -
ACCORDING TO AVAILABLE INFORMATION, THIS RIGHT VENTRICULAR LEAD AND DEVICE REMAIN IMPLANTED AND IN SERVICE. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS EVENT WILL BE UPDATED.
ADDITIONAL INFORMATION WAS RECEIVED. A FURTHER OUT OF RANGE SHOCK IMPEDANCE MEASUREMENT HAD BEEN RECORDED, HOWEVER THERE HAD BEEN NO ALERT, HOWEVER TS REVIEWED THE DATA AND DETERMINED THE OUT OF RANGE MEASUREMENT HAD OCCURRED PRIOR TO THE RESET OF PREVIOUS OUT OF RANGE MEASUREMENT. FURTHER REVIEW OF THE DATA DETERMINED THAT THE SHOCK IMPEDANCE MEASUREMENT REVEALED NOISE DURING DEVICE PACING. FURTHER LEAD INTEGRITY INVESTIGATION WAS RECOMMENDED.
A FOLLOW UP VISIT WAS PERFORMED. NO NOISE EPISODES WERE OBSERVED DURING POCKET MANIPULATION, ISOMETRIC AND ARM MOVEMENTS. SHOCK IMPEDANCE MEASUREMENTS WERE NORMAL WITH THE LEFT ARM IN NORMAL POSITION AND HIGH WHEN THE LEFT ARM WAS POSITIONED ABOVE THE HEAD. SEVERAL MANEUVERS REVEALED SIMILAR RESULTS. IT WAS THOUGHT THERE WAS AN ISSUE AT THE DEVICE POCKET SITE CAUSING AN INTERMITTENT CONTACT WITH THE DEVICE. ALTHOUGH LEAD INTEGRITY TESTING WAS RECOMMENDED, DUE TO THE PATIENT¿S CONDITION, A DECISION WAS MADE TO FURTHER MONITOR THIS SYSTEM WITH REMOTE MONITORING AND FREQUENT FOLLOW UP VISITS. ALL RV LEAD MEASUREMENTS WERE NORMAL. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A FOLLOW UP VISIT, A SAVE TO DISK WAS PROVIDED TO TECHNICAL SERVICES FOR THEIR REVIEW. SEVERAL HIGH OUT OF RANGE SHOCK LEAD IMPEDANCE MEASUREMENTS HAVE BEEN RECORDED AND THE RIGHT VENTRICULAR LEAD IMPEDANCE MEASUREMENTS HAVE DECREASED DURING THE PAST FEW MONTHS. THE ARRHYTHMIA LOGBOOK REVEALED NUMEROUS VENTRICULAR EPISODES HAD OCCURRED AND THERE IS NOISE ON THE RIGHT VENTRICULAR AND SHOCK CHANNELS. IN ADDITION ANTITACHYCARDIA PACING WAS DELIVERED. THE NOISE COULD NOT BE REPRODUCED BY THE PHYSICIAN. THE DEVICE WAS REPROGRAMMED. FURTHER SYSTEM INTEGRITY INCLUDING AN HIGH RESOLUTION X-RAY OF THE HEADER WAS RECOMMENDED TO VERIFY WHETHER THERE MAY BE A SUB OPTIMAL CONNECTION. NO ADVERSE PATIENT EFFECTS WERE REPORTED. A DECISION HAS BEEN MADE TO PERFORM A LEAD REVISION IN THE NEAR FUTURE.
- -
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142654 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0181 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 0181| 0171| 4480| 4554| P107 |