FDA Adverse Event
Injury
Summary report: N
EASYTRAK 2
MDR report key: 3041164
·
Received April 8, 2013
Report
- Report Number
- 2124215-2013-04878
- Event Type
- Injury
- Date Received
- April 8, 2013
- Date of Event
- January 10, 2013
- Report Date
- January 10, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- OJX
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IA
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
Additional Manufacturer Narrative · 1
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.
Description of Event or Problem · 1
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD HAD DISLODGED. THE PHYSICIAN EXPLANTED THIS LEAD AND REPLACED WITH A DIFFERENT LEAD. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143408 | EASYTRAK 2 | IMPLANTABLE LEAD | OJX | CPI - DEL CARIBE | 4542 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 55 YR | Hospitalization| L| R | 0293| N140| 5072| 4542| 4591| 4554 |