FDA Adverse Event Injury Summary report: N

EASYTRAK 2

MDR report key: 3041164 · Received April 8, 2013

Report

Report Number
2124215-2013-04878
Event Type
Injury
Date Received
April 8, 2013
Date of Event
January 10, 2013
Report Date
January 10, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
OJX
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
IA
Reporter Occupation
MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A THOROUGH EVALUATION OF THE LEAD WAS PERFORMED. RESISTANCE AND PRESSURE TESTS WERE COMPLETED TO ASSESS LEAD ELECTRICAL PERFORMANCE AND INSULATION INTEGRITY. MEASUREMENTS THROUGHOUT THESE TESTS WERE WITHIN NORMAL LIMITS. MICROSCOPIC INSPECTIONS OF THE TERMINAL PIN ASSEMBLY, LEAD BODY, AND ELECTRODE TIP FOUND NO ANOMALIES. LABORATORY TESTING WAS UNABLE TO REPRODUCE THE REPORTED CLINICAL OBSERVATIONS AND DETAILED ANALYSIS DID NOT REVEAL ANY ABNORMALITIES.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS LEAD HAD DISLODGED. THE PHYSICIAN EXPLANTED THIS LEAD AND REPLACED WITH A DIFFERENT LEAD. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143408 EASYTRAK 2 IMPLANTABLE LEAD OJX CPI - DEL CARIBE 4542

Patients

Seq Age Sex Outcome Treatment
1 55 YR Hospitalization| L| R 0293| N140| 5072| 4542| 4591| 4554