VITALITY
Report
- Report Number
- 2124215-2013-00679
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- January 10, 2013
- Report Date
- January 11, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI) AFTER EXPERIENCING TWO CHARGE TIMES GREATER THAN THE CHARGE TIME LIMIT. END OF LIFE (EOL) WAS DECLARED FOLLOWING A SINGLE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE AND THE RESULTS INDICATED THE MONITORING VOLTAGE WAS NORMAL BASED ON THERAPY USE AND PROGRAMMED SETTINGS. IT WAS CONCLUDED THAT THIS DEVICE DID NOT EXPERIENCE PREMATURE BATTERY DEPLETION. RATHER, ERI WAS REACHED EARLIER THAN EXPECTED DUE TO A HIGHER-THAN-TYPICAL BUILDUP OF INTERNAL BATTERY IMPEDANCE. THE DEVICE WAS CAPABLE OF DETECTING AND TREATING ARRHYTHMIAS, AS THE BATTERY ITSELF HAD SUFFICIENT CAPACITY REMAINING TO PROVIDE THERAPY IF NEEDED.
THE DEVICE IS CURRENTLY UNDERGOING ANALYSIS. WHEN ADDITIONAL INFORMATION BECOMES AVAILABLE, THIS REPORT WILL BE UPDATED.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS PRODUCT WAS RETURNED WITH NO REPORTED PRODUCT PERFORMANCE ISSUES AND NO REPORTED ADVERSE PATIENT EFFECTS. INITIAL ANALYSIS COMPLETED IN OUR POST MARKET QUALITY ASSURANCE LABORATORY IDENTIFIED A PRODUCT PERFORMANCE ISSUE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143314 | VITALITY | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T125 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 53 YR | T125| 0185| E143| 4470 |