ENDOTAK RELIANCE
Report
- Report Number
- 2124215-2013-01699
- Event Type
- Injury
- Date Received
- April 8, 2013
- Date of Event
- January 16, 2013
- Report Date
- January 30, 2013
- Manufacturer
- CPI - DEL CARIBE
- Product Code
- NVY
- PMA / PMN Number
- P910073
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MO
- Reporter Occupation
- PHYSICIAN
Narratives
THE PRODUCT HAS BEEN RECEIVED FOR ANALYSIS. THIS REPORT WILL BE UPDATED UPON COMPLETION OF ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, IT WAS NOTED THAT THE LEAD WAS RETURNED SEVERED AT 68 MM FROM THE TERMINAL PIN WITH ONLY THE PROXIMAL END OF THE IS-1 TERMINAL LEG BEING RETURNED. BLOOD AND BODY FLUID WERE OBSERVED ON THE TERMINAL PIN AND TERMINAL MOLDING AND THE INSULATION WAS FOUND TO BE CUT AND TORN AROUND THE CIRCUMFERENCE 24 MM FROM THE TERMINAL PIN. FURTHER VISUAL INSPECTION FOUND A SEPARATION IN THE INSULATION AND STRETCHED CONDUCTOR COILS FROM APPROXIMATELY 20 TO 60 MM FROM THE TERMINAL PIN. ANALYSIS DETERMINED THAT ALL DAMAGE APPEARED TO BE RELATED TO THE EXPLANT PROCEDURE. THE RETURNED LEAD SEGMENT WAS SUBJECT TO DIRECT CURRENT RESISTANCE TESTING AND PASSED ON ALL PATHS, VERIFYING THIS PORTION OF THE LEAD'S ELECTRICAL PERFORMANCE WAS INTACT. THE LABORATORY ANALYSIS SPECULATED THAT IT IS POSSIBLE THAT BLOOD AND BODY FLUID IF IT INGRESSES INTO THE DEVICE PORT WHERE THE TERMINAL PIN IS LOCATED, COULD ACT LIKE A GLUE AND CAUSE THE TERMINAL END TO BE STUCK INSIDE THE DEVICE HEADER. HOWEVER, ANALYSIS WAS NOT ABLE TO CONFIRM THE FIELD ALLEGATION.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT DURING A DEVICE CHANGE OUT PROCEDURE FOR NORMAL BATTERY DEPLETION, THE RIGHT VENTRICULAR (RV) LEAD AS WELL AS THE COMPETITIVE RIGHT ATRIAL (RA) LEAD WERE NOTED TO BE STUCK IN THE HEADER OF THE DEVICE. THE PHYSICIAN ENSURED THAT ALL EIGHT SETSCREWS WERE LOOSENED WITH NO SUCCESS. WHEN REMOVING THE COMPETITIVE RA LEAD THE PHYSICIAN TUGGED SO HARD THAT THE TERMINAL PIN BECAME DETACHED FROM THE LEAD AND REMAINED IN THE HEADER. IN ORDER TO REMOVE THE RV LEAD, THE PHYSICIAN CUT IT OUT OF THE DEVICE HEADER. BOTH LEADS WERE SURGICALLY ABANDONED AND THEN EXPLANTED THE FOLLOWING DAY. NO ADDITIONAL ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143001 | ENDOTAK RELIANCE | IMPLANTABLE LEAD | NVY | CPI - DEL CARIBE | 0157 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Hospitalization| L| R | 1298| H179| 1823| 4513| 0157| 1488TC |