FDA Adverse Event Injury Summary report: N

ENDOTAK RELIANCE

MDR report key: 3041074 · Received April 8, 2013

Report

Report Number
2124215-2013-01202
Event Type
Injury
Date Received
April 8, 2013
Date of Event
January 11, 2013
Report Date
January 11, 2013
Manufacturer
CPI - DEL CARIBE
Product Code
NVY
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

AS NO FURTHER INFORMATION CONCERNING THIS REPORT IS EXPECTED, OUR INVESTIGATION IS COMPLETE. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS RIGHT VENTRICULAR LEAD WITH NON-BOSTON SCIENTIFIC DEVICE DISPLAYED NOISE AND OVERSENSING. THERE WAS CONCERN THAT THIS LEAD WAS FRACTURED. A REVISION PROCEDURE WAS PERFORMED. THIS LEAD WAS SURGICALLY ABANDONED AND REPLACED. NO ADVERSE PATIENT EFFECTS WERE REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143027 ENDOTAK RELIANCE IMPLANTABLE LEAD NVY CPI - DEL CARIBE 0165

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization| L| R