FDA Adverse Event Malfunction Summary report: N

ANIMAS INSULIN CARTRIDGE

MDR report key: 3040948 · Received April 8, 2013

Report

Report Number
2531779-2013-03780
Event Type
Malfunction
Date Received
April 8, 2013
Report Date
March 14, 2013
Manufacturer
ANIMAS CORPORATION
Product Code
LZG
PMA / PMN Number
K080639
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
CA, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

FOLLOW-UP # 1 DATE OF SUBMISSION 06/11/2013 - DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: DURING INVESTIGATION, THE DISPLAY WAS FOUND TO BE DIM AND DISCOLORED. THE DISPLAY WAS REPLACED WITH A TEST SCREEN FOR TESTING AND WAS FOUND TO FUNCTION PROPERLY. UNRELATED TO THE DISPLAY ISSUE, EVALUATION REVEALED THAT THE CARTRIDGE COMPARTMENT WAS CRACKED AT THE THREADS.

Additional Manufacturer Narrative · 1

THE PUMP HAS NOT BEEN RETURNED TO ANIMAS. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. IF THE DEVICE IS RETURNED, AN EVALUATION SHALL BE COMPLETED AND A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSIONS CAN BE MADE AT THIS TIME.

Description of Event or Problem · 1

ON (B)(6) 2013, THE REPORTER CONTACTED ANIMAS ALLEGING THAT THE DISPLAY SCREEN WAS DIM AND FADING. THE REPORTER NOTED EVIDENCE OF MOISTURE BEHIND THE DISPLAY SCREEN. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE DEVICE CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143232 ANIMAS INSULIN CARTRIDGE INSULIN INFUSION PUMP LZG ANIMAS CORPORATION

Patients

Seq Age Sex Outcome Treatment
1 15 YR