FDA Adverse Event Malfunction Summary report: N

LIGACLIP ENDOSCOPIC ROTATING CLIP APPLIER

MDR report key: 3040929 · Received April 8, 2013

Report

Report Number
3005075853-2013-01635
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 10, 2013
Report Date
March 13, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K864102
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). INFORMATION UNAVAILABLE. THE ANALYSIS RESULTS FOUND THAT THE DEVICE WAS RETURNED IN GOOD VISUAL CONDITION AND WITH ONE CLIP LOADED IN THE JAW. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, RETAINED AND FORMED THE REMAINING CLIPS AS INTENDED. THE DEVICE WAS FOUND TO BE FULLY FUNCTIONAL AND CONFORMING TO OUR MANUFACTURING SPECIFICATIONS. NO CONCLUSION COULD BE REACHED AS TO WHAT MAY HAVE CAUSED THE REPORTED INCIDENT. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, AN UNEXPECTED NOISE WAS HEARD WHEN THE CLIP WAS FED INTO THE JAWS. BESIDES, THE CLIP WAS NOT FORMED PROPERLY. THE DEVICE WAS FIRED OUT OF THE PATIENT AFTER THIS INCIDENT. ANOTHER DEVICE WAS USED TO COMPLETE THE CASE. THERE WERE NO ADVERSE CONSEQUENCES TO THE PATIENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
142877 LIGACLIP ENDOSCOPIC ROTATING CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1