ONETOUCHPING GLUCOSEMGMTSYSTEM
Report
- Report Number
- 2531779-2013-03778
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Report Date
- March 11, 2013
- Manufacturer
- ANIMAS CORPORATION
- Product Code
- LZG
- PMA / PMN Number
- K080639
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
THE PUMP HAS BEEN RETURNED TO ANIMAS. EVALUATION HAS NOT YET BEEN COMPLETED. ANIMAS HAS CONDUCTED A REVIEW OF THE DEVICE HISTORY RECORD FOR THIS PUMP AND CONFIRMED THAT IT WAS OPERATING WITHIN REQUIRED SPECIFICATIONS AT THE TIME OF RELEASE. WHEN EVALUATION IS COMPLETE A SUPPLEMENTAL REPORT WILL BE FILED. NO CONCLUSION CAN BE MADE AT THIS TIME.
FOLLOW-UP #1 (B)(4)-DEVICE EVALUATION: THE PUMP HAS BEEN RETURNED AND EVALUATED BY PRODUCT ANALYSIS ON (B)(4) 2013 WITH THE FOLLOWING FINDINGS: INVESTIGATION REVEALED THAT THE DISPLAY IS DIM; THE LETTERS HAVE A RED HUE. SETTING THE CONTRAST TO THE HIGHEST SETTING DID NOT IMPROVE THE DISPLAY. WHEN A TEST DISPLAY SCREEN WAS USED THE DISPLAY FUNCTIONED PROPERLY. DISPLAY LENS IS SCRATCHED. NO ACTIVITY OBSERVED IN PUMP HISTORIES RELATED TO THE COMPLAINT. UNRELATED TO THE COMPLAINT, THE KEYPAD SYMBOLS WERE WORN.
THE REPORTER CONTACTED ANIMAS ON (B)(6) 2013 AND STATED THE DISPLAY SCREEN WAS DIM/FADED, SCRATCHED, AND CRACKED AROUND THE EDGES. THIS COMPLAINT IS BEING REPORTED BECAUSE THE REPORTED ISSUE WAS NOT RESOLVED WITH TROUBLESHOOTING. THERE WAS NO INDICATION THAT THE PRODUCT CAUSED OR CONTRIBUTED TO AN ADVERSE EVENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143921 | ONETOUCHPING GLUCOSEMGMTSYSTEM | INSULIN INFUSION PUMP | LZG | ANIMAS CORPORATION |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 17 YR |