LIGACLIP ENDOSCOPIC ROTATING CLIP APPLIER
Report
- Report Number
- 3005075853-2013-01634
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Report Date
- March 14, 2013
- Manufacturer
- ETHICON ENDO-SURGERY, LLC.
- Product Code
- FZP
- PMA / PMN Number
- K864102
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- TN, US
- Reporter Occupation
- OTHER
Narratives
(B)(4). INFORMATION ANTICIPATED, BUT UNAVAILABLE AT THIS TIME.
(B)(4). THE DEVICE WAS RETURNED FOR ANALYSIS AND UPON INSPECTION THE JAWS WERE FOUND TO BE IN A YIELDED CONDITION MAKING THE DEVICE NON-FUNCTIONAL. POSSIBLE CAUSES FOR THE CONDITION FOUND MAY BE IF THE DEVICE IS CLOSED OVER AN EXISTING HARD OBJECT OR CLIP PLACING STRESS ON THE JAWS CAUSING THEM TO DISTORT OR YIELD AND NOT RETURN TO THEIR ORIGINAL DIMENSIONS/POSITION OR EXCESSIVE APPLICATION OF TORQUE TO THE JAWS WHEN POSITIONING THE DEVICE ON A VESSEL. NO INCIDENT RELATED TO THE REPORTED EVENT WAS OBSERVED DURING THE BATCH RECORD REVIEW.
IT WAS REPORTED THAT DURING AN UNKNOWN PROCEDURE, THE FIRST CLIP WAS FIRED AND IT DID NOT LOOK RIGHT; THE DEVICE WOULD NOT FIRE ANY FURTHER CLIPS. SECOND DEVICE WAS OPENED AND THE CASE WAS COMPLETED NO PATIENT CONSEQUENCE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142892 | LIGACLIP ENDOSCOPIC ROTATING CLIP APPLIER | CLIP, IMPLANTABLE | FZP | ETHICON ENDO-SURGERY, LLC. | UNK | UNK |