FDA Adverse Event Malfunction Summary report: N

GYNECARE X-TRACT MORCELLATOR

MDR report key: 3040892 · Received April 8, 2013

Report

Report Number
2210968-2013-03582
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 20, 2013
Report Date
March 20, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K993801
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MI, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). ADDITIONAL NARRATIVE: THE PROCEDURE WAS COMPLETED WITH ANOTHER LIKE DEVICE.

Additional Manufacturer Narrative · 1

(B)(4). CONCLUSION: THE ACTUAL DEVICE INVOLVED IN THIS EVENT WAS RETURNED FOR EVALUATION. DURING THE EVALUATION THE CPC CONNECTOR WAS FOUND TO BE MISALIGNED.

Additional Manufacturer Narrative · 1

(B)(4) - DAMAGE TO DRIVE CONNECTOR OR COUPLING. CONCLUSION: THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED HAS BEEN RECEIVED, HOWEVER, THE PRODUCT EVALUATION IS NOT YET COMPLETE. ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM. IN ADDITION, A REVIEW OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED AND THE DEVICE MET ALL FINISHED GOODS RELEASE CRITERIA.

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2013. DURING THE PROCEDURE, IT WAS DIFFICULT TO ATTACH THE DISPOSABLE TO THE CONNECTOR PORT ON THE DEVICE. IT WAS NOT REPORTED HOW THE CASE WAS COMPLETED. THERE WERE NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143854 GYNECARE X-TRACT MORCELLATOR LAPAROSCOPIC MORCELLATOR HET ETHICON, INC. NA NA

Patients

Seq Age Sex Outcome Treatment
1