FDA Adverse Event Malfunction Summary report: N

LIGAMAX-5MM ENDO CLIP APPLIER

MDR report key: 3040891 · Received April 8, 2013

Report

Report Number
3005075853-2013-01632
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
February 27, 2013
Report Date
March 12, 2013
Manufacturer
ETHICON ENDO-SURGERY, LLC.
Product Code
FZP
PMA / PMN Number
K050344
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TN, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

(B)(4). BATCH# K9028E, EXPIRATION DATE: 12/16/2017, MANUFACTURING DATE: 1/16/2013. TWO DEVICE WERE RETURNED: DEVICE (A) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING CLIPS AS INTENDED. IN ORDER TO CONFIRM THE CLIPS WERE WITHIN MANUFACTURING SPECIFICATIONS, THE CLIPS WERE EVALUATED USING A TOOL THAT IS DESIGNED TO DETERMINE PROPER FORMATION. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. DEVICE (B) WAS RETURNED IN GOOD VISUAL CONDITION. IN AN ATTEMPT TO REPLICATE THE REPORTED INCIDENT, THE DEVICE WAS TESTED FOR FUNCTIONALITY. UPON TESTING, THE DEVICE WAS CYCLED, FED, AND FORMED THE REMAINING ELEVEN CLIPS AS INTENDED. IN ORDER TO CONFIRM THE CLIPS WERE WITHIN MANUFACTURING SPECIFICATIONS, THE CLIPS WERE EVALUATED USING A TOOL THAT IS DESIGNED TO DETERMINE PROPER FORMATION. DURING ANALYSIS THE JAWS OPEN AND CLOSE AS INTENDED. IN ADDITION THE DEVICE LOCKED OUT AS INTENDED. THE MANUFACTURING RECORDS WERE REVIEWED AND NO DISCREPANCIES WERE FOUND DURING THE MANUFACTURING PROCESS.

Description of Event or Problem · 1

IT WAS REPORTED THAT DURING A LAPAROSCOPIC CHOLECYSTECTOMY PROCEDURE, THE SURGEON REPORTED THAT THE FIRST DEVICE WORKED FOR A FEW CLIPS AND THEN MADE A LOUD CRUNCH NOISE. THE CLIP WOULD NOT CLOSE AROUND THE VESSEL. A SECOND DEVICE WAS OPENED AND THE DEVICE EITHER JAMMED OR DID NOT DEPLOY THE CLIP. THERE WERE NO PATIENT CONSEQUENCES. CASE WAS COMPLETED WITH SUTURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143630 LIGAMAX-5MM ENDO CLIP APPLIER CLIP, IMPLANTABLE FZP ETHICON ENDO-SURGERY, LLC. UNK UNK

Patients

Seq Age Sex Outcome Treatment
1