FDA Adverse Event Malfunction Summary report: N

VITALITY 2

MDR report key: 3040833 · Received April 8, 2013

Report

Report Number
2124215-2013-00176
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
October 4, 2010
Report Date
January 4, 2013
Manufacturer
GUIDANT CRM CLONMEL IRELAND
Product Code
LWS
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
HEALTH PROFESSIONAL

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A LONGEVITY CALCULATION CONFIRMED THE DEVICE DID NOT MEET LONGEVITY ESTIMATES PROVIDED IN DEVICE LABELING. PORTIONS OF THE CIRCUITRY, INCLUDING THE BATTERY, WERE ISOLATED AND TESTED. FINAL ANALYSIS CONCLUDED THE PREMATURE BATTERY DEPLETION WAS DUE TO LOW-VOLTAGE CAPACITOR DEGRADATION, WHICH RESULTED IN A HIGH CURRENT CONDITION. (B)(6) REMAINS IMPLANTED/INCREASED DEVICE FOLLOW-UP ADVISED: TECHNICAL SERVICES ADVISED THAT THE DEVICE FOLLOW-UP INTENSIFY TO MONTHLY. NO ADVERSE PATIENT EFFECTS HAVE BEEN REPORTED. AT THIS TIME, RECORDS INDICATE THAT THIS DEVICE REMAINS IMPLANTED. THIS INVESTIGATION WILL BE UPDATED SHOULD FURTHER INFORMATION BE PROVIDED.

Additional Manufacturer Narrative · 1

UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE DID NOT MEET A PORTION OF THE INITIAL TESTING. ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.

Description of Event or Problem · 1

BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE DECLARED ELECTIVE REPLACEMENT INDICATOR (ERI), WITH A CHARGE TIME MEASUREMENT OF 21 SECONDS AND A MONITORING VOLTAGE MEASUREMENT OF 2.42 V. IT WAS DISCUSSED THAT ERI LIKELY DECLARED AS A RESULT OF CHARGE TIME MEASUREMENT. APPROXIMATELY TWO YEARS LATER, THE DEVICE WAS EXPLANTED FOR NORMAL BATTERY DEPLETION. NO ADVERSE PATIENT EFFECTS WERE REPORTED DURING THE PROCEDURE. THE DEVICE WAS THEN RETURNED TO ALLOW FOR RELIABILITY ANALYSIS.