LUMAX 340 DR-T
Report
- Report Number
- 1028232-2013-00924
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 6, 2013
- Report Date
- March 25, 2013
- Manufacturer
- BIOTRONIK SE & CO. KG
- Product Code
- MRM
- PMA / PMN Number
- P000009
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- OR, US
- Reporter Occupation
- NURSE
Narratives
UPON RECEIPT, THE DEVICE INTERROGATION REVEALED THE BATTERY STATUS ERI. A NUMBER OF 24 CHARGING CYCLES WAS DOCUMENTED. THE ABILITY OF THE DEVICE TO DELIVER THERAPIES WAS VERIFIED. THE ANTI-BRADYCARDIA PACING PULSES PROVED TO BE FLAWLESS IN AMPLITUDE AND FREQUENCY AS PROGRAMMED. A FIBRILLATION SIGNAL WAS APPLIED AND THE DEVICE DELIVERED A DEFIBRILLATION SHOCK AS SPECIFIED, DOCUMENTING A CORRECT SENSING AND SHOCK DELIVERY. IN PARTICULAR, THE SPECIFIED ENERGY LEVEL WAS REACHED. THE ICD WAS IMPLANTED FOR 43 MONTHS; 24 CHARGING CYCLES WERE DOCUMENTED IN THE ICD'S MEMORY. THE AMOUNT OF CHARGE TAKEN FROM THE BATTERY WAS VERIFIED. THE BATTERY CONDITION WAS FOUND TO BE ANTICIPATED. IN SUMMARY, THE ICD WAS FULLY FUNCTIONAL. THE ERI BATTERY STATUS WAS ANTICIPATED.
THIS DEVICE IS AT ERI INDICATION AND WAS REPLACED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 144164 | LUMAX 340 DR-T | ICD | MRM | BIOTRONIK SE & CO. KG | 355267 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 52 YR | Hospitalization |