LUMAX 340 DR-T

MDR report key: 3040788 · Received April 8, 2013

Report

Report Number
1028232-2013-00924
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 6, 2013
Report Date
March 25, 2013
Manufacturer
BIOTRONIK SE & CO. KG
Product Code
MRM
PMA / PMN Number
P000009
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
OR, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

UPON RECEIPT, THE DEVICE INTERROGATION REVEALED THE BATTERY STATUS ERI. A NUMBER OF 24 CHARGING CYCLES WAS DOCUMENTED. THE ABILITY OF THE DEVICE TO DELIVER THERAPIES WAS VERIFIED. THE ANTI-BRADYCARDIA PACING PULSES PROVED TO BE FLAWLESS IN AMPLITUDE AND FREQUENCY AS PROGRAMMED. A FIBRILLATION SIGNAL WAS APPLIED AND THE DEVICE DELIVERED A DEFIBRILLATION SHOCK AS SPECIFIED, DOCUMENTING A CORRECT SENSING AND SHOCK DELIVERY. IN PARTICULAR, THE SPECIFIED ENERGY LEVEL WAS REACHED. THE ICD WAS IMPLANTED FOR 43 MONTHS; 24 CHARGING CYCLES WERE DOCUMENTED IN THE ICD'S MEMORY. THE AMOUNT OF CHARGE TAKEN FROM THE BATTERY WAS VERIFIED. THE BATTERY CONDITION WAS FOUND TO BE ANTICIPATED. IN SUMMARY, THE ICD WAS FULLY FUNCTIONAL. THE ERI BATTERY STATUS WAS ANTICIPATED.

Description of Event or Problem · 1

THIS DEVICE IS AT ERI INDICATION AND WAS REPLACED. SHOULD ADDITIONAL INFORMATION BECOME AVAILABLE, THIS FILE WILL BE UPDATED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
144164 LUMAX 340 DR-T ICD MRM BIOTRONIK SE & CO. KG 355267

Patients

Seq Age Sex Outcome Treatment
1 52 YR Hospitalization