VITALITY 2
Report
- Report Number
- 2124215-2013-00147
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- January 3, 2013
- Report Date
- January 15, 2013
- Manufacturer
- GUIDANT CRM CLONMEL IRELAND
- Product Code
- LWS
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- OTHER
Narratives
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, THE DEVICE DID NOT MEET A PORTION OF INITIAL TESTING. ANALYSIS OF THE RETURNED PRODUCT IS CURRENTLY ONGOING. THIS EVENT WILL BE UPDATED UPON COMPLETION OF THE ANALYSIS.
UPON RECEIPT AT OUR POST MARKET QUALITY ASSURANCE LABORATORY, A REVIEW OF DEVICE MEMORY REVEALED THAT THE DEVICE DECLARED END OF LIFE (EOL) AFTER EXPERIENCING ONE CHARGE TIME GREATER THAN 30 SECONDS. THE OBSERVED RATE OF BATTERY USAGE WAS COMPARED TO THE EXPECTED RATE OF BATTERY USAGE; THE RESULTS INDICATED THAT THE MONITORING VOLTAGE WAS NORMAL. ALTHOUGH THE BATTERY ITSELF HAD NOT DEPLETED PREMATURELY, EOL WAS TRIGGERED EARLIER THAN EXPECTED BY EXTENDED CHARGE TIMES, AND THE ERI TO EOL TIME PERIOD WAS SHORTENED, DUE TO A HIGHER-THAN-TYPICAL BUILD-UP OF INTERNAL BATTERY IMPEDANCE.
BOSTON SCIENTIFIC RECEIVED INFORMATION THAT THIS DEVICE WAS RETURNED TO ALLOW FOR RELIABILITY ANALYSIS. NO ADVERSE PATIENT EFFECTS WERE REPORTED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 143731 | VITALITY 2 | IMPLANTABLE CARDIOVERTER DEFIBRILLATOR | LWS | GUIDANT CRM CLONMEL IRELAND | T177 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 76 YR | E141| T177| 0174| 0184 |