FDA Adverse Event Malfunction Summary report: N

PRIME 5TH WHEEL STRETCHER, 26"

MDR report key: 3040605 · Received April 8, 2013

Report

Report Number
0001831750-2013-02990
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 15, 2013
Report Date
March 15, 2013
Manufacturer
STRYKER MEDICAL-KALAMAZOO
Product Code
FPO
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NY, US
Reporter Occupation
OTHER

Narratives

Description of Event or Problem · 1

IT WAS REPORTED BY SERVICE REPORT THAT THE BRAKES WOULD NOT ENGAGE DUE BOTH HEAD AND FOOT END BRAKE PEDALS BEING DISCONNECTED FROM THE BRAKE ROD. ADDITIONALLY, THE MATTRESS COULD NOT ADHERE TO THE LITTER SURFACE DUE TO THE VELCRO PILES MISSING. NO PATIENT WAS AFFECTED AND NO ADVERSE CONSEQUENCE OR CLINICALLY RELEVANT DELAY IN TREATMENT WAS REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143674 PRIME 5TH WHEEL STRETCHER, 26" STRETCHER, WHEELED FPO STRYKER MEDICAL-KALAMAZOO 1105000026