FDA Adverse Event Malfunction Summary report: N

PERMOBIL M300 CORPUS 3G

MDR report key: 3040591 · Received March 12, 2013

Report

Report Number
1221084-2013-00006
Event Type
Malfunction
Date Received
March 12, 2013
Date of Event
February 13, 2013
Report Date
March 12, 2013
Manufacturer
PERMOBIL, INC.
Product Code
ITI
PMA / PMN Number
K103477
Product Problem
Yes
Report Source
Manufacturer report
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

WHEELCHAIR WAS INSPECTED BY MFR AT THE USER'S HOME. BASED ON A VISUAL INSPECTION, THE WIRES ATTACHED TO THE SWITCHBOX WAS DAMAGED AFTER THE CLIENT RAN INTO SOMETHING. THE DAMAGE TO THE WIRES CAUSED THE WIRES TO SHORT CIRCUIT. THE SWITCHBOX HAS BEEN REMOVED AND THE CLIENT IS COMFORTABLE OPERATING THE SEATING FUNCTIONS THROUGH THE JOYSTICK.

Description of Event or Problem · 1

THE SWITCHBOX WAS MOUNTED AT THE END OF THE ARMREST OPPOSITE THE JOYSTICK ON THE POWER WHEELCHAIR. THE MOUNTING BRACKET HAD BROKEN LIKELY AS A RESULT OF THE USER RUNNING INTO SOMETHING, WHICH CAUSED STRESS AT THE POINT WHERE THE WIRE EXITS THE SWITCHBOX. THE WIRE HAD A SHORT CIRCUIT AND CAUSED SOME SMOKE. THE INSULATION OF THE WIRES IS SELF EXTINGUISHING. THE USER DID NOT REPORT ANY INJURY. THE SWITCHBOX WAS REMOVED FROM THE WHEELCHAIR AND THE WHEELCHAIR FUNCTIONS PROPERLY WITHOUT THE SWITCHBOX ATTACHED. THE USER IS COMFORTABLE OPERATING THE SEAT FUNCTIONS THROUGH THE JOYSTICK.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104493 PERMOBIL M300 CORPUS 3G POWER WHEELCHAIR ITI PERMOBIL, INC.

Patients

Seq Age Sex Outcome Treatment
1