POSEY TWICE-AS-TOUGH WRIST CUFF, QUICK RELEASE CUFF
Report
- Report Number
- 2020362-2013-00110
- Event Type
- Malfunction
- Date Received
- March 12, 2013
- Report Date
- February 15, 2013
- Manufacturer
- J. T. POSEY CO.
- Product Code
- FMQ
- PMA / PMN Number
- K963413
- Removal / Correction Number
- NA
- Report Source
- Manufacturer report
- Reporter Location
- IL, US
- Reporter Occupation
- OTHER
Narratives
RESULTS - EVAL OF THE RETURNED UNIT CONFIRMED DAMAGE TO THE FEMALE SIDE OF THE QUICK RELEASE BUCKLE. THE BUCKLE IS CRACKED BUT MALE SIDE WILL CLICK INTO FEMALE SIDE. CONNECTING STRAP BUCKLE CLOSES WITH OR WITHOUT STRAP FED THROUGH AND LOCKS SECURELY. CONNECTING STRAP BUCKLE UNLOCKS AS IT SHOULD. NOTE: INSTRUCTIONS FOR USE STATE: BEFORE EACH USE, CHECK CUFFS AND STRAPS FOR CRACKS, TEARS, AND/OR EXCESSIVE WEAR OR STRETCH; CRACKED OR BROKEN BUCKLES OR LOCKS; AND/OR THAT HOOK AND LOOP ADHERES SECURELY, AS THESE MAY ALLOW PATIENT TO REMOVE CUFF. (B)(4).
CUSTOMER REPORTED AFTER THE RESTRAINT HAD BEEN LAUNDERED, THEY NOTICED THERE WAS DAMAGE TO THE FEMALE SIDE OF THE QUICK RELEASE BUCKLE. CUSTOMER CONFIRMS THE RECOMMENDED LAUNDERING INSTRUCTIONS HAVE BEEN FOLLOWED. THIS WAS DISCOVERED DURING SET UP AND THERE WAS NO PATIENT INVOLVEMENT. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 104324 | POSEY TWICE-AS-TOUGH WRIST CUFF, QUICK RELEASE CUFF | FMQ | J. T. POSEY CO. | 279Q | 120710 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |