FDA Adverse Event Malfunction Summary report: N

POSEY TWICE-AS-TOUGH WRIST CUFF, QUICK RELEASE CUFF

MDR report key: 3040575 · Received March 12, 2013

Report

Report Number
2020362-2013-00110
Event Type
Malfunction
Date Received
March 12, 2013
Report Date
February 15, 2013
Manufacturer
J. T. POSEY CO.
Product Code
FMQ
PMA / PMN Number
K963413
Removal / Correction Number
NA
Report Source
Manufacturer report
Reporter Location
IL, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

RESULTS - EVAL OF THE RETURNED UNIT CONFIRMED DAMAGE TO THE FEMALE SIDE OF THE QUICK RELEASE BUCKLE. THE BUCKLE IS CRACKED BUT MALE SIDE WILL CLICK INTO FEMALE SIDE. CONNECTING STRAP BUCKLE CLOSES WITH OR WITHOUT STRAP FED THROUGH AND LOCKS SECURELY. CONNECTING STRAP BUCKLE UNLOCKS AS IT SHOULD. NOTE: INSTRUCTIONS FOR USE STATE: BEFORE EACH USE, CHECK CUFFS AND STRAPS FOR CRACKS, TEARS, AND/OR EXCESSIVE WEAR OR STRETCH; CRACKED OR BROKEN BUCKLES OR LOCKS; AND/OR THAT HOOK AND LOOP ADHERES SECURELY, AS THESE MAY ALLOW PATIENT TO REMOVE CUFF. (B)(4).

Description of Event or Problem · 1

CUSTOMER REPORTED AFTER THE RESTRAINT HAD BEEN LAUNDERED, THEY NOTICED THERE WAS DAMAGE TO THE FEMALE SIDE OF THE QUICK RELEASE BUCKLE. CUSTOMER CONFIRMS THE RECOMMENDED LAUNDERING INSTRUCTIONS HAVE BEEN FOLLOWED. THIS WAS DISCOVERED DURING SET UP AND THERE WAS NO PATIENT INVOLVEMENT. THE CUSTOMER DID NOT PROVIDE A DATE WHEN THIS WAS DISCOVERED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
104324 POSEY TWICE-AS-TOUGH WRIST CUFF, QUICK RELEASE CUFF FMQ J. T. POSEY CO. 279Q 120710

Patients

Seq Age Sex Outcome Treatment
1 NA