FDA Adverse Event Malfunction Summary report: N

CAREASSIST BED

MDR report key: 3040482 · Received March 13, 2013

Report

Report Number
9615739-2013-00067
Event Type
Malfunction
Date Received
March 13, 2013
Date of Event
February 13, 2013
Report Date
February 13, 2013
Manufacturer
HILL-ROM PLUVIGNER
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
TX, US
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

THE TECHNICIAN INSTALLED NEW BRAKE CASTERS AT THE HEAD END OF THE BED TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE TECHNICIAN ALLEGED THE BRAKE CASTERS SET BUT ARE NOT HOLDING AT THE HEAD END OF THE BED. NO PT IMPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
106234 CAREASSIST BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM PLUVIGNER 1170

Patients

Seq Age Sex Outcome Treatment
1