FDA Adverse Event Malfunction Summary report: N

STELLARIS 23GA POSTERIOR PACK

MDR report key: 3040480 · Received April 3, 2013

Report

Report Number
1920664-2013-00054
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 8, 2013
Report Date
March 8, 2013
Manufacturer
BAUSCH & LOMB, INC.
Product Code
HQC
PMA / PMN Number
K101325
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
KS
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

EVALUATION COMPLETED. ONE OPENED BL5623 POUCH FROM LOT U9434 WAS RETURNED. THE TIP PROTECTOR IS IN PLACE. THE NEEDLE IS NOT BENT. THE PORT WINDOW IS IN THE OPENED POSITION. THERE IS DRIED SOLUTION VISIBLE IN THE ASPIRATION LINE. THE BACK CAP IS ALIGNED CORRECTLY WITH THE VENT HOLE IN THE SIDE OF THE CUTTER BODY. A FUNCTIONAL TEST WAS PERFORMED USING A STELLARIS PC SYSTEM. THE INNER NEEDLE WILL NOT MOVE. THE CUTTER DOES ASPIRATE BUT WILL NOT CUT. THE STERILIZATION AND LOT HISTORY RECORDS HAVE BEEN REVIEWED AND FOUND TO BE ACCEPTABLE.

Description of Event or Problem · 1

THE USER FACILITY IN (B)(6) REPORTED THE CUTTER WORKED CORRECTLY DURING TRIAL CONNECTION. THEY STARTED WITH A LOW CUTTER SPEED; HOWEVER WHEN GEARED UP TO HIGHER SPEED THE CUTTER FUNCTION WAS NOT WORKING PROPERLY. DURING SURGERY THE CUTTER FAILED TO PROVIDE ADEQUATE CUTTING ACTION, BUT ASPIRATION CONTINUED. THERE WAS NO IMPACT TO THE PATIENT REPORTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136902 STELLARIS 23GA POSTERIOR PACK HQC BAUSCH & LOMB, INC. BL5623 U9434

Patients

Seq Age Sex Outcome Treatment
1