LIFEVEST WCD 4000 SYSTEM
Report
- Report Number
- 3008642652-2013-00730
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 28, 2013
- Report Date
- March 12, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- BIOMEDICAL ENGINEER
Narratives
DEVICE EVAL OF MONITOR (B)(4) HAS BEEN COMPLETED. UPON INVESTIGATION THE MONITOR WAS RESETTING. THE CAUSE FOR THE RESETS WAS ISOLATED TO AN INTERMITTENT BGA SOLDER CONNECTION AT THE U500 DSP COMPONENT ON THE MONITOR C/A BOARD. THE ROOT CAUSE FOR THE INTERMITTENT BGA CONNECTION AT U500 COULD NOT BE POSITIVELY IDENTIFIED, BUT THE DAMAGE LIKELY RESULTED FROM THE MONITOR IMPACTING A HARD SURFACE. ROOT CAUSE ANALYSIS OF SIMILAR EVENTS FOUND THAT EXCESSIVE STRAIN ON THE PRINTED CIRCUIT ASSEMBLY (PCA) FROM SEVERE IMPACT CAN CAUSE FRACTURED SOLDER CONNECTIONS. A MECHANICAL DESIGN CHANGE TO REDUCE THE PCA STRAIN WAS APPROVED BY FDA ON (B)(4) 2012. IMPLEMENTATION BEGAN (B)(4) 2013. RESULTS WILL BE MONITORED. NO ADVERSE EVENT RESULTED FROM THE DEFECTIVE MONITOR. THE LAST PT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
A REVIEW OF SERVICE DATA DETECTED A REPORTABLE PROBLEM. DURING SERVICING, MONITOR (B)(4) WAS RESETTING. THE LAST PT TO USE THIS MONITOR DID NOT REPORT ANY DEFICIENCIES.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106262 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | NA |