FDA Adverse Event
Malfunction
Summary report: N
LIFEVEST WCD 4000 SYSTEM
MDR report key: 3040466
·
Received March 13, 2013
Report
- Report Number
- 3008642652-2013-00716
- Event Type
- Malfunction
- Date Received
- March 13, 2013
- Date of Event
- February 12, 2013
- Report Date
- March 12, 2013
- Manufacturer
- ZOLL LIFECOR CORPORATION
- Product Code
- MVK
- PMA / PMN Number
- P010030
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- NOT APPLICABLE
Narratives
Additional Manufacturer Narrative · 1
DEVICE EVAL SUMMARY - DEVICE EVAL OF MONITOR SN (B)(4) HAS BEEN COMPLETED. THE REPORTED PROBLEM (WILL NOT POWER UP) HAS BEEN CONFIRMED. UPON INVESTIGATION THE PLD (PROGRAMMABLE LOGIC DEVICE) U1003 WAS SHORTED AT THE 1.8V LINE ON THE MONITOR C/A BOARD, WHICH PREVENTED THE MONITOR FROM POWERING UP. THE ROOT CAUSE FOR THE SHORTED U1003 COMPONENT COULD NOT BE POSITIVELY IDENTIFIED. NO ADVERSE EVENT RESULTED FROM THE SHORTED PLD. THE PT RECEIVED A REPLACEMENT MONITOR.
Description of Event or Problem · 1
A (B)(6) MALE PT CALLED ZOLL CUSTOMER SUPPORT TO REPORT THAT HIS MONITOR WOULDN'T POWER UP. THE PT WAS ISSUED A REPLACEMENT MONITOR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 106177 | LIFEVEST WCD 4000 SYSTEM | WEARABLE CARDIOVERTER DEFIBRILLATOR | MVK | ZOLL LIFECOR CORPORATION | WCD 4000 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 61 YR |