HEARTSINE SAMARITAN 300P AND PAD PAK
Report
- Report Number
- 3004123209-2012-00492
- Event Type
- Malfunction
- Date Received
- February 11, 2013
- Date of Event
- December 5, 2012
- Report Date
- January 20, 2013
- Manufacturer
- HEARTSINE TECHNOLOGIES, LTD
- Product Code
- MKJ
- PMA / PMN Number
- K052465
- Removal / Correction Number
- Z-0124-2013
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- SW
- Reporter Occupation
- NOT APPLICABLE
Narratives
ROUTINE TESTING CONFIRMED THAT THIS DEVICE WAS INSTALLED BY THE CUSTOMER IN (B)(6) 2010, AND THAT IT HAD PERFORMED TO SPECIFICATION UP TO (B)(6) 2012. THE INFORMATION OBTAINED FROM THE DEVICE SHOWED THAT THE DEVICE FAILED THE WEEKLY SELF-RUN TEST ON (B)(6) 2012, BECAUSE OF A LOW BATTERY. INVESTIGATION DID NOT CONFIRM A FAULT WITH THE DEVICE OR ANY COMPONENT IN THE DEVICE. DURING TESTING, THE DEVICE WAS SHOWN TO PERFORM TO SPECIFICATION. THE IDENTIFICATION OF THE LOW BATTERY FAULT MAY HAVE BEEN CAUSED BY EXTERNAL INFLUENCES SUCH THAT THE PAD-PAK MAY NOT HAVE BEEN SEATED CORRECTLY IN THE DEVICE. THERE MAY ALSO BE AN INTERMITTENT FAILURE OF THE BATTERY TERMINAL POGO PINS, ALTHOUGH INVESTIGATION COULD NOT REPLICATE A FAULT WITH THE POGO PINS. INVESTIGATION ALSO CONFIRMED THAT THE RETURNED PAD-PAKS FROM LOT 687 (USE UNTIL 07/2013) SHOWED SIGNS OF NORMAL USAGE AND THESE TOO PERFORMED T SPECIFICATION DURING TESTING. THE PAD PAK, WHICH CONTAINS THE ELECTRODES AND BATTERIES, IS LABELLED FOR SINGLE USE, BUT THE SAMARITAN PAD 300 AND 300P DEVICES ARE FOR MULTI-USE.
THERE WAS NO PT INVOLVED IN THIS EVENT. THIS DEVICE MALFUNCTIONED BECAUSE THE STATUS INDICATOR WAS FLASHING RED AND THE DEVICE WAS EMITTING A LOW BATTERY WARNING. A DEVICE IN THIS FAULT MODE, IF LEFT UNDETECTED, COULD RESULT IN THE FAILURE OF THE DEVICE TO PERFORM AS INTENDED, IF REQUIRED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 60490 | HEARTSINE SAMARITAN 300P AND PAD PAK | AUTOMATED EXTERNAL DEFIBRILLATOR | MKJ | HEARTSINE TECHNOLOGIES, LTD | PAD |