INFINITI VISION SYSTEM OZIL

MDR report key: 3040431 · Received April 3, 2013

Report

Report Number
2028159-2013-00605
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
March 7, 2013
Report Date
March 7, 2013
Manufacturer
ALCON-IRVINE TECHNOLOGY CENTER
Product Code
HQC
PMA / PMN Number
K082845
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NJ, US
Reporter Occupation
NURSE

Narratives

Additional Manufacturer Narrative · 1

THE COMPANY REPRESENTATIVE EXAMINED THE SYSTEM. THE SYSTEM WAS TESTED FOR PROPER IRRIGATION AND ASPIRATION FUNCTIONS. THE REPORTED EVENT OF LACK OF IRRIGATION WAS NOT REPLICATED. UNRELATED TO THE REPORTED EVENT, A PREVENTIVE MAINTENANCE (PM) WAS ALSO PERFORMED. THE SYSTEM WAS THEN TESTED AND MET ALL PRODUCT SPECIFICATIONS. THERE WAS NO SAMPLE RETURNED FOR EVALUATION AND NO ADDITIONAL INFORMATION PROVIDED RELATED TO THIS EVENT. FOR THIS REASON, STEPS COULD NOT BE TAKEN TO REPLICATE OR CONFIRM THE REPORTED EVENT. A REVIEW OF THE STM'S EVENT LOG SHOWED NO EVIDENCE TO CONFIRM THE REPORTED EVENT. A REVIEW OF COMPLAINTS FOR THE LAST 24 MONTHS DID NOT INDICATE ANY ADDITIONAL RELATED REPORTS FOR THIS SYSTEM. THE ROOT CAUSE COULD NOT BE DETERMINED. (B)(4).

Description of Event or Problem · 1

A NURSE REPORTED THAT THERE WAS NO FLUID WHEN IN QUADRANT MODE DURING A CATARACT WITHIN INTRAOCULAR LENS IMPLANT PROCEDURE. THE CASE WAS COMPLETED WITH NO PATIENT HARM. ADDITIONAL INFORMATION HAS BEEN REQUESTED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136654 INFINITI VISION SYSTEM OZIL PHACOFRAGMENTATION SYSTEM HQC ALCON-IRVINE TECHNOLOGY CENTER INFINITI OZIL NA