FDA Adverse Event Malfunction Summary report: N

GYNECARE MORCELLEX* TISSUE MORCELLATOR

MDR report key: 3040419 · Received April 8, 2013

Report

Report Number
2210968-2013-03574
Event Type
Malfunction
Date Received
April 8, 2013
Date of Event
March 13, 2013
Report Date
March 21, 2013
Manufacturer
ETHICON, INC.
Product Code
HET
PMA / PMN Number
K100280
Removal / Correction Number
NA
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
NL
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

(B)(4). THE PRODUCT UPON WHICH THIS MEDWATCH IS BASED IS ANTICIPATED. ONCE THE PRODUCT IS RECEIVED, ANY FURTHER INFORMATION DERIVED FROM THE EVALUATION WILL BE SUBMITTED IN A SUPPLEMENTAL 3500A FORM.

Additional Manufacturer Narrative · 1

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

Description of Event or Problem · 1

IT WAS REPORTED THAT A PATIENT UNDERWENT A GYNECOLOGICAL PROCEDURE ON (B)(6) 2013. DURING THE PROCEDURE, THE DEVICE STARTED TO WORK WORSE. THE KNIFE DIDN'T WORK WELL ANYMORE. ANOTHER LIKE DEVICE WAS USED TO COMPLETE THE PROCEDURE WITH NO ADVERSE PATIENT CONSEQUENCES.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
143453 GYNECARE MORCELLEX* TISSUE MORCELLATOR LAPAROSCOPE, GYNECOLOGIC HET ETHICON, INC. NA MT216731

Patients

Seq Age Sex Outcome Treatment
1