ACCU-CHEK SPIRIT COMBO
Report
- Report Number
- 2183996-2013-00608
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- April 2, 2013
- Report Date
- May 15, 2013
- Manufacturer
- ROCHE DIABETES CARE AG
- Product Code
- LZG
- PMA / PMN Number
- NA
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NL
- Reporter Occupation
- MEDICAL EQUIPMENT COMPANY TECHNICIAN/REPRESENTATIVE
Narratives
THE INCIDENT OCCURRED OUTSIDE THE UNITED STATES. INFORMATION CONTAINED WITHIN THIS REPORT IS ALL THAT IS AVAILABLE AT THIS TIME. IF FURTHER INFORMATION IS OBTAINED, IT WILL BE PROVIDED IN THE SUPPLEMENTAL REPORT.
THE CHECK BUTTON DOES NOT RESPOND. THERE ARE PARTICLES INSIDE THE HOUSING OF THE CHECK BUTTON. THE PARTICLES ISOLATE THE AREA OF CONTACT INSIDE THE BUTTON HOUSING. DUE TO THIS PROBLEM THE CHECK BUTTON DOES NOT RESPOND AND IS WITHOUT FUNCTION.
ON (B)(6) 2013, IT WAS REPORTED THAT CHECK BUTTON ON THE PATIENT'S INFUSION DEVICE WAS NOT FUNCTIONING. THE PROBLEM WAS NOTICED WHILE THE PATIENT WAS BEING TRAINED ON THE DEVICE. NO PHYSIOLOGICAL EFFECTS WERE REPORTED. THE PATIENT DID NOT REQUIRE MEDICAL ASSISTANCE FROM A HEALTHCARE PROFESSIONAL OR SECOND PARTY TO ADDRESS THE ISSUE. INFUSION DEVICE REPLACED AND WAS REQUESTED TO BE RETURNED FOR PRODUCT EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142561 | ACCU-CHEK SPIRIT COMBO | INSULIN INFUSION DEVICE | LZG | ROCHE DIABETES CARE AG | 00700006863 | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |