FDA Adverse Event
Malfunction
Summary report: N
CONTOUR
MDR report key: 3040396
·
Received April 8, 2013
Report
- Report Number
- 1826988-2013-00147
- Event Type
- Malfunction
- Date Received
- April 8, 2013
- Date of Event
- March 1, 2013
- Report Date
- March 16, 2013
- Manufacturer
- BAYER HEALTHCARE LLC
- Product Code
- NBW
- PMA / PMN Number
- K062058
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER
Narratives
Additional Manufacturer Narrative · 1
CORRECTION OF THE BLOOD RESULTS IN THE DESCRIPTION OF THE PROBLEM: THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 281MG/DL ON THE CONTOUR METER, RE-TESTED ON A DIFFERENT METER AND THE READING WAS 140MG/DL.
Description of Event or Problem · 1
THE CUSTOMER RECEIVED A BLOOD GLUCOSE READING OF 172MG/DL ON THE CONTOUR METER, RE-TESTED ON A DIFFERENT METER AND THE READING WAS 70MG/DL. THE DIFFERENCE BETWEEN THE READINGS FALLS IN THE "C" ZONE OF THE CONSENSUS ERROR GRID, MAKING THE DIFFERENCE CLINICALLY SIGNIFICANT NO ADVERSE EVENT WAS ALLEGED. CONTROL SOLUTION WAS SENT TO THE CUSTOMER FOR FURTHER TROUBLESHOOTING. PRODUCT IS NOT EXPECTED TO BE RETURNED FOR EVALUATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 142559 | CONTOUR | BLOOD GLUCOSE TEST STRIPS | NBW | BAYER HEALTHCARE LLC | 7080G | 2JC3B01 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |