FDA Adverse Event Malfunction Summary report: N

840 VENTILATOR

MDR report key: 3040379 · Received April 3, 2013

Report

Report Number
8020893-2013-00788
Event Type
Malfunction
Date Received
April 3, 2013
Date of Event
February 18, 2013
Report Date
March 6, 2013
Manufacturer
COVIDIEN, FORMERLY NELLCOR PURITAN
Product Code
CBK
PMA / PMN Number
K970460
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
JA
Reporter Occupation
NOT APPLICABLE

Narratives

Description of Event or Problem · 1

COVIDIEN RECEIVED INFO STATING THAT DUE TO A MALFUNCTION OF AN 840 VENTILATOR THE PT WAS PLACED ON A SECOND VENTILATOR. THE PT WAS NOT HARMED OR INJURED AS A RESULT OF THE EVENT. THE COVIDIEN CUSTOMER SUPPORT ENGINEER (CSE) VERIFIED THE MALFUNCTION AND REPLACED THE OXYGEN SENSOR. THE UNIT PASSED EXTENDED SELF-TESTING.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
136324 840 VENTILATOR VENTILATOR CBK COVIDIEN, FORMERLY NELLCOR PURITAN 840

Patients

Seq Age Sex Outcome Treatment
1