AMIS
Report
- Report Number
- 3005180920-2008-00002
- Event Type
- Malfunction
- Date Received
- June 3, 2008
- Date of Event
- May 15, 2008
- Report Date
- May 15, 2008
- Manufacturer
- MEDACTA INTERNATIONAL, SA
- Product Code
- JEA
- PMA / PMN Number
- EXEMPT
- Removal / Correction Number
- NOT YET RECEIVED
- Adverse Event
- Yes
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- NOT APPLICABLE
Narratives
THE FIRST CLAMP ACCIDENTALLY OPENED RESULTING IN EXCESSIVE STRESSES ON THE SECOND CLAMP THAT SUBSEQUENTLY FRACTURED. IT IS HYPOTHESIZED THAT THE FIRST CLAMP MAY HAVE OPENED DUE TO INCORRECT SETTING OF THE ADJUSTING SCREW RESULTING IN INSUFFICIENT THREAD ENGAGEMENT. SAFETY NOTIFICATION ISSUED WITH UPDATED LABELING TO ASSURE THAT BOTH CLAMPS AND STRAPS ARE USED AND TO REINFORCE PROPER ADJUSTMENT AND CONNECTION OF CLAMPS.
A PRIMARY HIP SURGERY WAS BEING PERFORMED VIA AN AMIS (ANTERIOR MINIMALLY INVASIVE SURGERY) APPROACH AND USING AN AMIS MOBILE LEG POSITIONER AND UNIVERSAL TABLE CONNECTOR. DURING SURGERY TWO PROXIMAL CLAMPS OF THE UNIVERSAL TABLE FOR LEG POSITIONER DISCONNECTED FROM THE SURGICAL TABLE RESULTING IN THE PT FALLING TO THE FLOOR. THE PT SUFFERED HYPEREXTENSION OF ONE KNEE. IN ADDITION, THE ANESTHESIA TUBING WAS PULLED FROM THE PT'S THROAT. SURGERY WAS COMPLETED WITHOUT (INCIDENT AFTER REPOSITIONING THE PT ON THE SURGICAL TABLE.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 1 | AMIS | UNIVERSAL TABLE FOR LEG POSITIONER | JEA | MEDACTA INTERNATIONAL, SA | NA | 075146 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 66 YR | Required Intervention | LOT 062301| AMIS MOBILE LEG POSITIONER: (B)(4), |