FDA Adverse Event Malfunction Summary report: N

AMIS

MDR report key: 3040347 · Received June 3, 2008

Report

Report Number
3005180920-2008-00002
Event Type
Malfunction
Date Received
June 3, 2008
Date of Event
May 15, 2008
Report Date
May 15, 2008
Manufacturer
MEDACTA INTERNATIONAL, SA
Product Code
JEA
PMA / PMN Number
EXEMPT
Removal / Correction Number
NOT YET RECEIVED
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
GM
Reporter Occupation
NOT APPLICABLE

Narratives

Additional Manufacturer Narrative · 1

THE FIRST CLAMP ACCIDENTALLY OPENED RESULTING IN EXCESSIVE STRESSES ON THE SECOND CLAMP THAT SUBSEQUENTLY FRACTURED. IT IS HYPOTHESIZED THAT THE FIRST CLAMP MAY HAVE OPENED DUE TO INCORRECT SETTING OF THE ADJUSTING SCREW RESULTING IN INSUFFICIENT THREAD ENGAGEMENT. SAFETY NOTIFICATION ISSUED WITH UPDATED LABELING TO ASSURE THAT BOTH CLAMPS AND STRAPS ARE USED AND TO REINFORCE PROPER ADJUSTMENT AND CONNECTION OF CLAMPS.

Description of Event or Problem · 1

A PRIMARY HIP SURGERY WAS BEING PERFORMED VIA AN AMIS (ANTERIOR MINIMALLY INVASIVE SURGERY) APPROACH AND USING AN AMIS MOBILE LEG POSITIONER AND UNIVERSAL TABLE CONNECTOR. DURING SURGERY TWO PROXIMAL CLAMPS OF THE UNIVERSAL TABLE FOR LEG POSITIONER DISCONNECTED FROM THE SURGICAL TABLE RESULTING IN THE PT FALLING TO THE FLOOR. THE PT SUFFERED HYPEREXTENSION OF ONE KNEE. IN ADDITION, THE ANESTHESIA TUBING WAS PULLED FROM THE PT'S THROAT. SURGERY WAS COMPLETED WITHOUT (INCIDENT AFTER REPOSITIONING THE PT ON THE SURGICAL TABLE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
1 AMIS UNIVERSAL TABLE FOR LEG POSITIONER JEA MEDACTA INTERNATIONAL, SA NA 075146

Patients

Seq Age Sex Outcome Treatment
1 66 YR Required Intervention LOT 062301| AMIS MOBILE LEG POSITIONER: (B)(4),