INVACARE MOBILAIRE
Report
- Report Number
- 2182215-2013-00001
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- December 20, 2012
- Report Date
- April 2, 2013
- Manufacturer
- GARDNER DENVER THOMAS INC.
- Product Code
- CAF
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
ADDITIONAL DEVICE MFR DATE: 04/2012. UNITS RETURNED TO (B)(4) FOR BEING NOISY, BUT STILL FUNCTIONAL. (B)(4) PRELIMINARY EVAL WAS THAT THE UNITS HAD BAD BEARINGS. (B)(4) RETURNED THE UNITS TO THE (B)(4) DISTRIBUTOR (B)(4) FOR FURTHER EVAL. (B)(4) CONFIRMED THAT THEY RECEIVED THE UNITS. IT WAS NOTED THAT THE UNITS WERE IN A VERY USED CONDITION AND THAT THE FANS WERE VERY DIRTY. NO ADDITIONAL IN-DEPTH EVALS WERE CONDUCTED AND THE UNITS WERE SCRAPPED. AS THE ORIGINAL MFR OF THIS PRODUCT, IT IS ALWAYS OUR GOAL TO ESTABLISH A ROOT CAUSE AND TAKE CORRECTIVE ACTION CONCERNING ALL CUSTOMER COMPLAINTS. IN THIS PARTICULAR SITUATION SINCE THE UNITS WERE NOT RETURNED TO GARDNER DENVER THOMAS FOR ANALYSIS WE OBVIOUSLY CANNOT DETERMINE A POSITIVE ROOT CAUSE TO THE PROBLEM AND NO PERMANENT CORRECT ACTIONS WILL BE IMPLEMENTED. HOWEVER, IN AN EFFORT TO BE PROACTIVE WE HAVE REVIEWED OUR CUSTOMER COMPLAINT HISTORY OF THIS MODEL AND HAVE FOUND NO COMPLAINTS INVOLVING BEARING FAILURES. WE ALSO REVIEWED CUSTOMER COMPLAINTS OF SIMILAR NON-MEDICAL UNITS. IN THIS REVIEW, WE FOUND ONLY TWO BEARING FAILURE ISSUES ONE INVOLVED BEARING CORROSION OF A PUMP USED IN POND AERATION. IN THIS SITUATION WE CHANGED TO CORROSION RESISTANCE BEINGS. THE OTHER BEARING FAILURE WAS CAUSED BY AN INCORRECTLY MACHINED BEARING BORE. AT THE TIME OF THIS FAILURE THE BEARING BORE MACHINING WAS CONDUCTED BY AN OFF SITE MACHINE SHOP. THE CORRECTIVE ACTION INVOLVING THIS FAILURE WAS TO BRING THIS MACHINING PROCESS BACK INTO OUR FACILITY. BASED ON THE HISTORY OF THE NON-MEDICAL UNIT, CUSTOMER COMPLAINTS WE WILL IMPLEMENT ONE INTERIM ACTION. ON THE NEXT BUILD OF THE MEDICAL UNIT IN QUESTION (MODEL IRC 607) WE WILL CONDUCT A CAPABILITY STUDY INVOLVING THE MACHINING PROCESS OF THE CONNECTING ROD AND HOUSING BEARING BORES.
PRIVATE LABEL PRIMARY DISTRIBUTOR ((B)(4)): INDICATED UNITS OVERHEATED. NO INJURIES WERE ALLEGED. INITIAL REPORTER - SECONDARY DISTRIBUTOR ((B)(4)): INDICATED THAT UNITS WERE RETURNED FOR BEING NOISY BUT THEY WERE STILL FUNCTIONAL. NO INJURY WAS ALLEGED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136194 | INVACARE MOBILAIRE | NEBULIZER PUMP (CAF) | CAF | GARDNER DENVER THOMAS INC. | IRC607 |