FDA Adverse Event
Malfunction
Summary report: N
LIBERTY CYCLER SET, SINGLE CONN/EXT DL
MDR report key: 3040273
·
Received April 3, 2013
Report
- Report Number
- 8030665-2013-00177
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 5, 2013
- Report Date
- March 5, 2013
- Manufacturer
- REYNOSA MANUFACTURING
- Product Code
- FKX
- PMA / PMN Number
- K043363
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Occupation
- PATIENT
Narratives
Additional Manufacturer Narrative · 1
THE DEVICE WAS NOT RETURNED TO THE MFR FOR PHYSICAL EVAL AND THE FAILURE MODE CANNOT BE CONFIRMED. HOWEVER, AN INVESTIGATION OF THE DEVICE MANUFACTURING RECORDS WAS CONDUCTED BY THE MFR. THERE WERE NO DEVIATIONS OR NONCONFORMANCES DURING THE MANUFACTURING PROCESS. IN ADDITION, THE BATCH RECORD REVIEW CONFIRMED THE LABELING, MATERIAL, AND PROCESS CONTROLS WERE WITHIN SPEC.
Description of Event or Problem · 1
A PERITONEAL DIALYSIS PT HAS REPORTED THAT DIALYSIS SOLUTION WAS LEAKING OUT OF THE CASSETTE AND INTO THE CYCLER. PT WAS IN FILL FOUR OF TREATMENT. UPON REMOVING, THE TUBING SET FROM THE CYCLER, FLUID WAS NOTICED LEAKING INTO THE CYCLER. THE ORIGIN OF THE LEAK COULD NOT BE IDENTIFIED, PT HAD NO ILL EFFECTS. SAMPLE HAS NOT BEEN RETURNED TO THE MFR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136232 | LIBERTY CYCLER SET, SINGLE CONN/EXT DL | LIBERTY DIALYSIS CYCLER TUBING | FKX | REYNOSA MANUFACTURING | 12SR08801 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | LIBERTY DIALYSIS CYCLER |