SYSMEX POCH-100I ANALYZER
Report
- Report Number
- 1422681-2013-00006
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- November 30, 2012
- Report Date
- November 30, 2012
- Manufacturer
- SYSMEX CORP
- Product Code
- GKL
- PMA / PMN Number
- K032677
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- WA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
SC(B)(4) INFORMED SYSMEX AMERICA, INC ON 03/04/2013, THAT THE INVESTIGATION CONFIRMED THERE WAS POTENTIAL TO OMIT A CHARACTER, WHICH COULD BE A DECIMAL, A NUMBER, OR PART OF THE GRAPHICS FROM THE PRINTOUT DUE TO A DEFECT IN THE PRINTED CIRCUIT BOARD (PCB) OF THE INTERNAL PRINTER. THIS OMISSION OF A CHARACTER IS LIMITED TO THE THERMAL PRINTOUT AND DOES NOT AFFECT THE LCD DISPLAY OR DATA OUTPUT OF RESULTS. THE PROBABILITY OF OCCURRENCE WAS ESTIMATED AS ONE TIME OUT OF 2000 MEASUREMENTS. THE PRODUCT PROBLEM WAS IDENTIFIED AS A DEFECT OF THE PRINTED CIRCUIT BOARD (PCB) OF THE INTERNAL PRINTER. A CORRECTION OF THE AFFECTED POCH-100I ANALYZERS WILL BE PERFORMED BY REPLACING THE PRINTER PCB. OPERATORS OF THE 121 AFFECTED DEVICES WILL BE INFORMED OF A TEMPORARY COUNTERMEASURE; TO VERIFY THE PRINTOUT AGAINST THE LCD DISPLAY, OR AGAINST THE DATA TRANSMITTED TO THE LAB INFO SYSTEM UNTIL THE PCB CAN BE REPLACED. THE CORRECTION BY PCB REPLACEMENT IS TARGETED TO BE COMPLETED IN THE NEXT 90 DAYS.
THE OPERATOR REPORTED THAT THE THERMAL PRINTOUT ON THEIR POCH-100I AUTOMATED HEMATOLOGY ANALYZER INTERMITTENTLY DISPLAYED A NEUTROPHIL PERCENTAGE (NEUT%) FROM THE WHITE BLOOD CELL (WBC) DIFFERENTIAL WITHOUT A DECIMAL POINT. THE VALUE WAS CORRECT, BUT LACKED THE DECIMAL CAUSING THE NEUT% TO BE A NONSENSICAL VALUE. THE PRINTOUTS WERE FROM THE QC MODE, AND NO IMPACT TO PT TREATMENT AS A RESULT OF THE MISSING DECIMAL POINT WAS ALLEGED. THE COMPLAINT WAS ESCALATED TO SYSMEX CORP (B)(4) (SCJ) FOR FURTHER INVESTIGATION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136569 | SYSMEX POCH-100I ANALYZER | AUTOMATED HEMATOLOGY ANALYZER | GKL | SYSMEX CORP | POCH-1000I | F1844 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |