FDA Adverse Event
Other
Summary report: N
BRACHYVISION
MDR report key: 3037693
·
Received March 28, 2013
Report
- Report Number
- 3003793371-2013-00002
- Event Type
- Other
- Date Received
- March 28, 2013
- Date of Event
- January 1, 2013
- Report Date
- March 28, 2013
- Manufacturer
- VARIAN MEDICAL SYSTEMS, INC.
- Product Code
- MUJ
- PMA / PMN Number
- K992762
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- GM
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
Additional Manufacturer Narrative · 1
THIS REPORT IS BASED ON INFORMATION RECEIVED FROM A THIRD PARTY OR DEVELOPED BY VARIAN MEDICAL SYSTEMS.
Description of Event or Problem · 1
THE CUSTOMER USED BRACHYVISION VERSION 7.3.10 O CREATE INTERSTITIAL TREATMENT PLANS FOR 7 PTS WITH DIFFERENT TREATMENT SITES (INCLUDING NASOPHARYNX AND BASE OF TONGUE). DUE TO A MISUNDERSTANDING OF THE FRACTIONATION CONCEPT OF BRACHYVISION ALL 7 PTS WERE UNDERDOSED. FOR EXAMPLE, THEY PLANNED FOR 10 FRACTIONS WITH A TOTAL DOSE OF 35 GY BUT DELIVERED ONLY 0.35 GY PER FRACTION.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 129616 | BRACHYVISION | SYSTEM, PLANNING, RADIATION THERAPY TRMT | MUJ | VARIAN MEDICAL SYSTEMS, INC. |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | Other |