FDA Adverse Event
Malfunction
Summary report: N
OAKWORKS CFPM400 FLUOROSCOPY TABLE
MDR report key: 3035897
·
Received December 20, 2012
Report
- Report Number
- 2529571-2012-00001
- Event Type
- Malfunction
- Date Received
- December 20, 2012
- Date of Event
- November 20, 2012
- Report Date
- December 20, 2012
- Manufacturer
- OAKWORKS, INC.
- Product Code
- IXR
- PMA / PMN Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- RISK MANAGER
Narratives
Additional Manufacturer Narrative · 1
FLUOROSCOPY TABLE ACTUATOR FAILED CAUSING TABLE TO TILT FORWARD EXCESSIVELY, CAUSING PATIENT TO SLIDE OFF TABLE TOP TO FLOOR, NO INJURY OR INTERVENTION REPORTED. NO OTHER FAILURES OF THIS TYPE IN HISTORY OF PRODUCT CANNOT DETERMINE IF PRODUCT WAS MISUSED.
Description of Event or Problem · 1
ACTUATOR FAILED, ALLOWING TABLE TO RAPIDLY TILT FORWARD, PATIENT SLID OFF TABLE TO FLOOR. NO INJURY OR INTERVENTION REPORTED.