FDA Adverse Event Injury Summary report: N

GORE EXCLUDER AAA ENDOPROSTHESIS

MDR report key: 3034615 · Received March 27, 2013

Report

Report Number
2017233-2013-00175
Event Type
Injury
Date Received
March 27, 2013
Date of Event
September 10, 2010
Report Date
February 28, 2013
Manufacturer
W.L. GORE & ASSOCIATES
Product Code
MIH
PMA / PMN Number
P020004
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
GA, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

REFER TO THE JOURNAL OF VASCULAR SURGERY "LATE GORE EXCLUDER ENDOPROSTHESIS FABRIC TEAR LEADING TO ABDOMINAL AORTIC ANEURYSM RUPTURE 5 YEARS AFTER INITIAL IMPLANT" JANUARY 2013 (VOL. 57, NUMBER 1: 221-224).

Description of Event or Problem · 1

AS REPORTED IN THE JOURNAL OF VASCULAR SURGERY "LATE GORE EXCLUDER ENDOPROSTHESIS FABRIC TEAR LEADING TO ABDOMINAL AORTIC ANEURYSM RUPTURE 5 YEARS AFTER INITIAL IMPLANT" JANUARY 2013 (VOL. 57, NUMBER 1: 221-224), THIS PATIENT UNDERWENT TREATMENT OF A 6.7 CM ABDOMINAL AORTIC ANEURYSM WITH FOUR GORE EXCLUDER AAA ENDOPROSTHESES ON FEBRUARY 4, 2005. ON SEPTEMBER 10, 2010, THE PATIENT PRESENTED WITH LOWER ABDOMINAL AND BACK PAIN, AND A CT SCAN SHOWED A RETROPERITONEAL HEMATOMA THAT REPORTEDLY COULD BE ASSOCIATED WITH AN ANEURYSM RUPTURE DUE TO A TYPE III ENDOLEAK. AFTER RESUSCITATION AND SUBSTITUTION OF BLOOD PRODUCTS, THE PATIENT WAS TAKEN TO SURGERY. INTRA-OPERATIVE AORTOGRAPHY REPORTEDLY SHOWED SIGNIFICANT CONTRAST EXTRAVASATION INTO THE ANEURISMAL SAC AT THE SITE OF THE RUPTURE. IT WAS REPORTED THE EXACT LOCATION OF THE TEAR WAS DIFFICULT TO IDENTIFY, BUT THE FABRIC TEAR WAS MOST LIKELY LOCATED AT THE LEVEL OF THE FLOW DIVIDER AT THE SIDE OF THE RIGHT LIMB. PROXIMAL ANGIOPLASTY WAS PERFORMED, AND THEN THREE ADDITIONAL DEVICES WERE IMPLANTED. REPEAT ANGIOGRAPHY SHOWED NO EVIDENCE OF ENDOLEAK WITH GOOD FLOW INTO THE RENAL ARTERIES. THE PATIENT TOLERATED THE PROCEDURE.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
127343 GORE EXCLUDER AAA ENDOPROSTHESIS MIH / SYSTEM, ENDOVASCULAR GRAFT, AORTIC ANEURYSM TREATMENT MIH W.L. GORE & ASSOCIATES 03089223

Patients

Seq Age Sex Outcome Treatment
1 71 YR Hospitalization| R WARFARIN