UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM
Report
- Report Number
- 2122870-2013-00340
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- February 7, 2013
- Report Date
- March 4, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- DHA
- PMA / PMN Number
- K023764
- Removal / Correction Number
- Z-1736-2011
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- PA, US
- Reporter Occupation
- OTHER HEALTH CARE PROFESSIONAL
Narratives
THE FIELD SERVICE ENGINEER (FSE) ASSESSED THE INSTRUMENT AND NO ISSUES WERE NOTED. THE SAMPLE PROBE CARRYOVER TEST PASSED AND HIGH SENSITIVITY SYSTEM CHECK PASSED WITHIN SPECIFICATION. QUALITY CONTROL (QC) WAS WITHIN THE LABORATORY'S ESTABLISHED LIMITS. THE INSTRUMENT PERFORMED WITHIN PUBLISHED PERFORMANCE SPECIFICATIONS. A DEFINITIVE CAUSE OF THE INCIDENT IS UNKNOWN. THE PATIENT'S SAMPLE WAS COLLECTED IN A 13X100 MM GREEN TOP TUBE AND CENTRIFUGED AT 3,000 RPM (ROTATIONS PER MINUTE) FOR SEVEN MINUTES. THE SAMPLE WAS ANALYZED ON THE DAY OF COLLECTION AND REFRIGERATED AFTER THE ANALYSIS. SAMPLE QUALITY APPEARED NORMAL. THE LABORATORY PERFORMS QUALITY CONTROL (QC) EVERY EIGHT HOURS. QC WAS WITHIN RANGE AT THE TIME OF THE EVENT. THE CUSTOMER REANALYZED RECENT AND ANY NEW POSITIVE RESULTS FOR ACCURACY VERIFICATION.
THE CUSTOMER REPORTED ERRONEOUSLY ELEVATED BETA HUMAN CHORIONIC GONADOTROPHIN (HCG) RESULT, FOR ONE PATIENT, INVOLVING THE UNICEL DXI 800 ACCESS IMMUNOASSAY SYSTEM UTILIZED WITH THE ACCESS TOTAL SSHCG ASSAY AND CALIBRATORS. SUBSEQUENT ANALYSES OF THE PATIENT'S SAMPLE, ON AN ALTERNATE UNICEL DXI 800 SYSTEM, RECOVERED LOWER RESULTS, WITHIN THE NORMAL REFERENCE RANGE. THE ERRONEOUS RESULT WAS RELEASED OUT OF THE LABORATORY AND QUESTIONED BY THE PHYSICIAN AS IT DID NOT MATCH THE PATIENT'S CLINICAL PRESENTATION. AN AMENDED REPORT WAS ISSUED. THE CUSTOMER COULD NOT CONFIRM IF THERE WAS ANY CHANGE TO PATIENT CARE. THERE HAS BEEN NO REPORT OF PATIENT INJURY OR CHANGE IN PATIENT TREATMENT ASSOCIATED WITH THIS EVENT. THE INITIAL SAMPLE WAS TESTED THROUGH THE AUTOMATION LINE, AND THE REPEAT SAMPLE WAS PROGRAMMED MANUALLY AND TESTED THROUGH THE SPECIMEN PROCESSING UNIT. A BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) WAS DISPATCHED TO EVALUATE THE INSTRUMENT.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136536 | UNICEL® DXI 800 ACCESS® IMMUNOASSAY SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | DHA | BECKMAN COULTER | NA | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 34 YR |