FDA Adverse Event Injury Summary report: N

PARADIGM INSULIN INFUSION PUMP

MDR report key: 3034036 · Received April 3, 2013

Report

Report Number
3004209178-2013-92292
Event Type
Injury
Date Received
April 3, 2013
Date of Event
March 30, 2013
Report Date
March 30, 2013
Manufacturer
MEDTRONIC PUERTO RICO OPERATIONS CO.
Product Code
LZG
PMA / PMN Number
K040676
Adverse Event
Yes
Report Source
Manufacturer report
Reporter Location
NZ
Reporter Occupation
OTHER

Narratives

Additional Manufacturer Narrative · 1

CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.PLEASE SEE MEDWATCH REPORT # 3004209178-2013-92293.

Additional Manufacturer Narrative · 1

A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.

Description of Event or Problem · 1

IT WAS REPORTED THAT THE CUSTOMER WAS HOSPITALIZED DUE TO HIGH BLOOD GLUCOSE OF 19.1MMOL/L. IT WAS STATED THAT THE CUSTOMER'S INSULIN PUMP ALARMED MOTOR ERROR DURING BOLUS DELIVERY. TROUBLESHOOTING WAS PERFORMED. THE INSULIN PUMP PASSED THE DISPLACEMENT AND SELF TEST PASSED. THE DRIVE SUPPORT CAP APPEARS NORMAL. IT WAS MENTIONED THAT INSULIN WAS COMING OUT OF THE BACK OF THE RESERVOIR. ADVISED THE CALLER THAT THE INSULIN PUMP WOULD BE REPLACED AND REVERT TO BACK UP PLAN. NO FURTHER INFORMATION WAS PROVIDED.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
135712 PARADIGM INSULIN INFUSION PUMP INSULIN INFUSION PUMP LZG MEDTRONIC PUERTO RICO OPERATIONS CO. MMT-715NAL

Patients

Seq Age Sex Outcome Treatment
1 Hospitalization