PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP
Report
- Report Number
- 3004209178-2013-92307
- Event Type
- Injury
- Date Received
- April 3, 2013
- Date of Event
- March 17, 2013
- Report Date
- March 22, 2013
- Manufacturer
- MEDTRONIC PUERTO RICO OPERATIONS CO.
- Product Code
- OYC
- PMA / PMN Number
- P980022
- Adverse Event
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- MA
- Reporter Occupation
- PATIENT
Narratives
CURRENTLY IT IS UNKNOWN WHETHER OR NOT THE DEVICE MAY HAVE CAUSED OR CONTRIBUTED TO THE EVENT AS NO PRODUCT HAS BEEN RETURNED. NO CONCLUSION CAN BE DRAWN AT THIS TIME. THE DEVICE WILL BE RETURNED FOR ANALYSIS AND FURTHER INFORMATION WILL FOLLOW ONCE THE ANALYSIS HAS BEEN COMPLETED.
A COMPLETE ANALYSIS AND TESTING OF THE INSULIN PUMP SHOWED THAT IT WAS FUNCTIONING PROPERLY AND PASSED ALL FUNCTIONAL TESTING. AFTER TESTING IT WAS CONCLUDED THAT THE DEVICE OPERATED WITHIN SPECIFICATIONS.
THE CUSTOMER REPORTED HAVING AN URGENT CARE VISIT DUE TO HIGH BLOOD GLUCOSE OF 331MG/DL. TROUBLESHOOTING WAS PERFORMED. THE CUSTOMER STATED THAT HIS INSULIN PUMP WENT THOUGH THE BODY SCANNER WHEN HE FLEW TO (B)(6). THE CUSTOMER STATED THAT HE HAD A PEANUT BUTTER AND JELLY SANDWICH THE NIGHT BEFORE, AND WHEN HE WOKE UP HIS BLOOD GLUCOSE WAS HIGH. THE TIME, DATE, BASAL RATES, AND BOLUS WIZARD WERE CORRECT. THE DRIVE SUPPORT CAP APPEARS NORMAL. ASSISTED THE CUSTOMER TO RUN A MANUAL PRIME TEST AND THE INSULIN DID EXIT. PERFORMED THE HIGH PRESSURE TEST AND PASSED. THE CUSTOMER DID NOT FEEL COMFORTABLE AND REQUESTED A REPLACEMENT OF THE DEVICE. NO FURTHER INFORMATION WAS PROVIDED.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 135688 | PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP | INSULIN INFUSION PUMP / SENSOR AUGMENTED | OYC | MEDTRONIC PUERTO RICO OPERATIONS CO. | MMT-523NAB |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 62 YR | Hospitalization |