UNICEL® DXC 660I SYNCHRON® ACCESS CLINICAL SYSTEM
Report
- Report Number
- 2050012-2013-00236
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- March 13, 2013
- Report Date
- March 13, 2013
- Manufacturer
- BECKMAN COULTER
- Product Code
- JJE
- PMA / PMN Number
- K060256
- Removal / Correction Number
- NA
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- NC, US
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4).
A CUSTOMER REPORTED TO BECKMAN COULTER (BEC) THAT THE UNICEL DXC 660I SYNCHRON ACCESS CLINICAL SYSTEM GENERATED QUALITY CONTROL RESULTS OUTSIDE THE LABORATORY'S ESTABLISHED RANGES. WHILE TROUBLESHOOTING, THE CUSTOMER FOUND THAT THE INSTRUMENT LEAKED FLUID ONTO THE DRIP TRAY ON TOP OF THE REAGENT CAROUSEL AND ON APPROXIMATELY 80% OF THE REAGENT CARTRIDGES IN THE REAGENT COMPARTMENT. THE LEAK WAS CONTAINED WITHIN THE INSTRUMENT. THE CUSTOMER WAS WEARING GLOVES AND A LAB COAT AT THE TIME OF THE INCIDENT. THERE WAS NO REPORT OF INJURY OR EXPOSURE TO OPEN WOUNDS OR MUCOUS MEMBRANES, AND MEDICAL ATTENTION WAS NOT SOUGHT. THE CUSTOMER DID NOT REVIEW THE MATERIAL SAFETY DATA SHEET (MSDS), BUT HAS AN EXPOSURE CONTROL PLAN IN PLACE AT THE FACILITY. THERE WAS NO IMPACT TO PATIENT RESULTS OR TREATMENT. BECKMAN COULTER FIELD SERVICE ENGINEER (FSE) FOUND THAT THE REAGENT PROBE B WAS LEAKING DEIONIZED WATER. THE FSE REVIEWED THE INSTRUMENT EVENT LOG AND FOUND THAT THE INSTRUMENT GENERATED SEVERAL ERROR MESSAGES. ALTHOUGH THE OPERATOR WAS ALERTED TO A PROBLEM VIA ERROR MESSAGES AND THE INSTRUMENT STOPPED GENERATING RESULTS, THE ERROR MESSAGES WERE FOR MOTION ERRORS AND WERE NOT RELATED TO THE LIQUID LEAKING FROM THE REAGENT PROBE B COLLAR WASH. THE ERROR MESSAGES WOULD NOT ALERT THE OPERATOR THAT LIQUID LEAKED INTO THE REAGENT COMPARTMENT, POTENTIALLY CONTAMINATING THE REAGENTS. UPON FURTHER INVESTIGATION THE FSE FOUND THAT THE SMART MODULE C3 WAS NOT FUNCTIONING PROPERLY. THE FSE NOTED THAT THE SMART MODULE C3 FAILURE RESULTED IN A CASCADE OF FAILURES, INCLUDING THE REAGENT PROBE B COLLAR WASH VALVE WHICH WAS STUCK AND NOT VACUUMING FLUID AWAY FROM REAGENT PROBE B. AS A RESULT, FLUID LEAKED INTO THE REAGENT COMPARTMENT AND ONTO MOST OF THE REAGENT CARTRIDGES. THE FSE REPLACED THE SMART MODULE C3, AND VERIFIED THE REPAIR.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 137229 | UNICEL® DXC 660I SYNCHRON® ACCESS CLINICAL SYSTEM | ANALYZER, CHEMISTRY (PHOTOMETRIC, DISCRETE), FOR CLINICAL USE | JJE | BECKMAN COULTER | DXC 660I | NA |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 |