FDA Adverse Event Injury Summary report: N

DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE

MDR report key: 3033669 · Received April 3, 2013

Report

Report Number
2182208-2013-00495
Event Type
Injury
Date Received
April 3, 2013
Date of Event
March 5, 2013
Report Date
March 5, 2013
Manufacturer
RICE CREEK MFG
Product Code
NVN
PMA / PMN Number
P950024
Adverse Event
Yes
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MN, US
Reporter Occupation
PHYSICIAN

Narratives

Additional Manufacturer Narrative · 1

THE INFORMATION SUBMITTED REFLECTS ALL RELEVANT DATA RECEIVED. IF ADDITIONAL RELEVANT INFORMATION IS RECEIVED, A SUPPLEMENTAL REPORT WILL BE SUBMITTED. (B)(4).

Description of Event or Problem · 1

IT WAS REPORTED THAT FAR FIELD R-WAVE (FFRW) OVERSENSING WAS NOTED ON A REMOTE TRANSMISSION. THE ASSOCIATED DEVICE WAS REPROGRAMMED TO REDUCE OR ELIMINATE THE FAR FIELD R-WAVE (FFRW) SENSING. THE LEAD REMAINS IN USE. NO PATIENT COMPLICATIONS HAVE BEEN REPORTED AS A RESULT OF THIS EVENT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
137110 DRUG ELUTING PERMANENT RIGHT VENTRICULAR (RV) OR RIGHT ATRIAL (RA) PACEMAKER ELE NVN RICE CREEK MFG 4965-25

Patients

Seq Age Sex Outcome Treatment
1 00011 YR Required Intervention