TREK CORONARY DILATATION CATHETER
Report
- Report Number
- 2024168-2013-01949
- Event Type
- Malfunction
- Date Received
- April 3, 2013
- Date of Event
- February 20, 2013
- Report Date
- February 21, 2013
- Manufacturer
- AV-TEMECULA-CT
- Product Code
- LOX
- PMA / PMN Number
- K103110
- Product Problem
- Yes
- Report Source
- Manufacturer report
- Reporter Location
- JO
- Reporter Occupation
- HEALTH PROFESSIONAL
Narratives
(B)(4). DURING PROCESSING OF THIS COMPLAINT, ATTEMPTS WERE MADE TO OBTAIN COMPLETE EVENT, PATIENT AND DEVICE INFORMATION. THE DEVICE WAS RECEIVED. INVESTIGATION IS NOT YET COMPLETE. A FOLLOW-UP REPORT WILL BE SUBMITTED WITH ALL ADDITIONAL RELEVANT INFORMATION.
(B)(4). EVALUATION SUMMARY: THE DEVICE WAS RETURNED FOR EVALUATION. THE REPORTED SHAFT SEPARATION WAS CONFIRMED. THE REPORTED FAILURE TO ADVANCE COULD NOT BE REPLICATED IN A TESTING ENVIRONMENT AS IT WAS BASED ON OPERATIONAL CIRCUMSTANCES. BASED ON ANALYSIS OF THE RETURNED DEVICE AND SCANNING ELECTRON MICROSCOPY ANALYSIS, THERE IS NO INDICATION OF A PRODUCT DEFICIENCY. A REVIEW OF THE LOT HISTORY RECORD REVEALED NO NON-CONFORMANCES FOR THIS LOT. THE RESULTS OF THE QUERY OF SIMILAR INCIDENTS IN THE COMPLAINT HANDLING DATABASE FOR THIS LOT DID NOT INDICATE A MANUFACTURING ISSUE. BASED ON THE INFORMATION REVIEWED, THERE IS NO EVIDENCE TO INDICATE THE PRESENCE OF A PRODUCT DEFICIENCY.
IT WAS REPORTED THAT THE PROCEDURE WAS TO TREAT A MILDLY TORTUOUS, HEAVILY CALCIFIED, AND 90% STENOSED MID LEFT ANTERIOR DESCENDING ARTERY. THE PROXIMAL SHAFT OF THE 25X12 MM TREK BALLOON CATHETER FRACTURED IN TWO PIECES DURING A FAILED ATTEMPT TO CROSS. THE DEVICE WAS REMOVED FROM THE PATIENT WITHOUT ISSUE AND A NEW TREK BALLOON WAS USED SUCCESSFULLY TO COMPLETE THE CASE. THERE WAS NO CLINICALLY SIGNIFICANT DELAY IN PROCEDURE AND NO ADVERSE PATIENT EFFECTS. NO ADDITIONAL INFORMATION WAS PROVIDED.
SUBSEQUENT TO THE PREVIOUSLY FILED MEDWATCH REPORT, NEW INFORMATION RECEIVED AS FOLLOWS: THE BALLOON CATHETER HAD BEEN USED PREVIOUSLY FOR INFLATION PRIOR TO THE FAILED ATTEMPT TO RECROSS.
Devices
| Seq | Brand | Generic | Product Code | Manufacturer | Model | Lot | UDI-DI |
|---|---|---|---|---|---|---|---|
| 136199 | TREK CORONARY DILATATION CATHETER | CORONARY DILATATION CATHETER | LOX | AV-TEMECULA-CT | 20719G3 |
Patients
| Seq | Age | Sex | Outcome | Treatment |
|---|---|---|---|---|
| 1 | 68 YR |