FDA Adverse Event Malfunction Summary report: N

ADVANTA BED

MDR report key: 3033348 · Received March 1, 2013

Report

Report Number
1824206-2013-01160
Event Type
Malfunction
Date Received
March 1, 2013
Date of Event
February 1, 2013
Report Date
February 1, 2013
Manufacturer
HILL-ROM, INC.
Product Code
FNL
Product Problem
Yes
Report Source
Manufacturer report
Reporter Location
MO, US
Reporter Occupation
UNKNOWN

Narratives

Additional Manufacturer Narrative · 1

THE ACCOUNT TOOK THE COMMUNICATION CABLE OFF THE BED AND FOUND A PIN HAD BEEN BROKEN OFF ON THE UNIVERSAL TV POWER CONTROL BOARD. THE ACCOUNT REPLACED THE UNIVERSAL TV POWER CONTROL BOARD TO RESOLVE THE ISSUE.

Description of Event or Problem · 1

THE ACCOUNT ALLEGED THE NURSE CALL IS NOT FUNCTIONING. NO PT INPACT.

Devices

Seq Brand Generic Product Code Manufacturer Model Lot UDI-DI
90302 ADVANTA BED A/C POWERED ADJUSTABLE HOSPITAL BED FNL HILL-ROM, INC. 1600

Patients

Seq Age Sex Outcome Treatment
1